Clinical apheresis · Practical reference

POCARD Plasma Adsorption
Prescription Guide

ImmunoCore® · Immuno-Adsopak® · ABO Adsopak® A / B / AB · Lipopak® · NucleoCore®
Separated plasma only · Device selection, technical planning, safety and printable order worksheet

Clinical-use limitation. Independent educational prescribing aid; not POCARD's instructions for use (IFU), a validated medical order, a CE/MDR certification statement or proof of French/European marketing authorization. Verify the precise model and size, country-specific authorization, current manufacturer IFU, plasma separator and machine compatibility, priming, pressure/flow limits, anticoagulation, regeneration and handling instructions before treatment. Product webpages and brochures are not sufficient to prescribe a device.
Route and reuse safety. All procedures in this document route cell-free separated plasma through an adsorber. Do not apply any of these instructions to a whole-blood hemoperfusion cartridge. Reusable POCARD products require patient-dedicated handling and exact validated regeneration/preservation steps; a manufacturer's stated maximum number of cycles is not permission to reuse a cartridge without its full IFU.

Scope, method and evidence boundaries

Plasma adsorption (PA) separates plasma from blood cells, directs plasma across a target-specific adsorbent and reinfuses the processed plasma with blood cells. The plasma is processed, not routinely discarded and replaced as in therapeutic plasma exchange (TPE). Therapy must be matched to the patient's disease, target molecule, technical constraints and local authorizations.

Separated plasma5 product familiesABO A / B / AB subtypesModel-specific reuseNo whole-blood prescriptions
Included product families: ImmunoCore (single-use broad Ig immunoadsorption); Immuno-Adsopak (regenerable broad Ig immunoadsorption); ABO Adsopak (anti-A, anti-B or dual-antibody adsorption); Lipopak (atherogenic lipoprotein adsorption); and NucleoCore (extracellular DNA/DNA-containing structure adsorption). All are identified by POCARD as plasma-route columns. [1–6]

Not included: non-adsorptive plasma separators, plasma-exchange replacement consumables and adsorbers whose current primary product labeling specifies direct whole-blood hemoperfusion rather than a documented plasma indication. No whole-blood treatment setpoints are provided here.

Evidence vs intended target: Manufacturer lists of diseases or molecular targets are not equivalent to independent efficacy evidence, guideline recommendations, or country-specific approved indications. The maturity of supporting evidence is particularly limited for emerging cfDNA/NETs adsorption. [2–8]

Plasma adsorber catalogue and clinical targets

Broad immunoadsorption

ImmunoCore®

Plasma immunoadsorber with an immobilized synthetic ligand binding immunoglobulins/autoantibodies and circulating immune complexes.

300 mL housing · single-use · [2,7]

Regenerable immunoadsorption

Immuno-Adsopak®

Plasma immunoadsorber for immunoglobulins, autoantibodies and immune complexes; reusable only within the stated same-patient protocol.

300 mL housing · manufacturer lists 5 uses · [3,7]

ABO-selective immunoadsorption

ABO Adsopak® A / B / AB

Synthetic A/B-antigen-mimetic ligands selectively bind anti-A, anti-B or both anti-A and anti-B isoagglutinins.

A, B, AB variants · size/use rules vary · [4,8]

Lipid apheresis

Lipopak®

Selective plasma adsorption of atherogenic lipoproteins, notably LDL and Lp(a), via an immobilized ligand.

300 mL housing · up to 18 regeneration cycles claimed · [5]

Emerging DNA-target adsorption

NucleoCore®

Designed to capture extracellular/cell-free DNA and associated DNA-containing structures, including NET-related materials, from plasma.

50 mL / 150 mL housing · single-use · [6]

Ordering principle

Exact model matters

Do not substitute a broad Ig sorbent for an antigen-specific ABO column, or a lipoprotein adsorber for an anti-DNA column. Ask the supplier for the exact product code, IFU revision and local regulatory documentation.

Country-specific market checks are mandatory

Plasma sorbentBinding targetPlasma-route clinical applicationReuse class / check
ImmunoCore®Ig / autoantibodies / circulating immune complexesSpecialist Ig immunoadsorption in antibody-mediated disease or transplant rejection protocolsSingle-use; check exact IFU [2]
Immuno-Adsopak®Ig / autoantibodies / immune complexesMulti-session or twin-column immunoadsorption (when specifically validated)Regenerable, same patient only; detailed handling per IFU [3,7]
ABO Adsopak® AAnti-A isoagglutininsABO-incompatible organ transplant pathway when anti-A must be reducedSpecific variant and size determine reuse [4,8]
ABO Adsopak® BAnti-B isoagglutininsABO-incompatible organ transplant pathway when anti-B must be reducedSpecific variant and size determine reuse [4,8]
ABO Adsopak® ABAnti-A and anti-B isoagglutininsWhen both isoagglutinins require removal in a documented protocolSpecific variant and size determine reuse [4,8]
Lipopak®LDL and Lp(a) / atherogenic lipoproteinsRefractory severe lipid disorders, specialist LDL/Lp(a) apheresisRegenerable; cycle number is not session number [5]
NucleoCore®Cell-free DNA and DNA-containing structures / NET-related materialInvestigational or locally authorized cfDNA-removal pathway; not a routine established disease indicationSingle-use; evidence and authorization review required [6]

ABO Adsopak sizes and antibody matching

Choose the A/B/AB ligand based on the recipient's clinically relevant antibody specificity against donor antigens, with laboratory transfusion/transplant review. A model letter describes the antibody removed, not simply the patient's blood group.

Model familyRemovesSizes identified in the manufacturer February 2026 brochureHandling information from that brochure
ABO Adsopak–AAnti-A antibodiesA50 · A150 · A30050 mL variant listed as single-use; 150 / 300 mL versions described as patient-specific reusable
ABO Adsopak–BAnti-B antibodiesB50 · B150 · B30050 mL variant listed as single-use; 150 / 300 mL versions described as patient-specific reusable
ABO Adsopak–ABAnti-A and anti-B antibodiesAB50 · AB30050 mL variant listed as single-use; 300 mL version described as patient-specific reusable
Variant listing discrepancy: the manufacturer's 2026 transplant brochure lists eight named ABO variants (including 50 mL forms), whereas the current product webpage summarizes 150 mL and 300 mL housings. Confirm that the exact SKU, market availability and reuse status remain current before ordering. The brochure's maximum 20 uses for larger units is neither a sterilization recipe nor a universal guarantee. [4,8]
Before selecting anti-A or anti-B

Confirm donor/recipient ABO types, baseline anti-A/anti-B IgG and IgM titers using a consistent laboratory method, desensitization drugs, organ-transplant protocol and the required target titer.

During a series of treatments

Retest clinically relevant isoagglutinin titers, assess rebound, coordinate transplantation timing and define thresholds for repeat adsorption, change of modality or stopping. No universal target titer applies.

Clinical applications and response measures

ImmunoCore / Immuno-Adsopak

POCARD describes immune-mediated neurologic disorders, systemic autoimmune disease and antibody-mediated graft rejection among potential use cases. Select indications only after specialist and guideline review; catalogue examples do not establish the same evidence category for each diagnosis.

Endpoints: organ function, symptoms or disease activity; disease-relevant antibodies (e.g., donor-specific antibodies) when interpretable; total Ig for safety. [2,3,7]

ABO Adsopak

Pre-/post-transplant ABO-incompatible organ pathways; other manufacturer-described ABO antibody indications require a separately justified clinical protocol.

Endpoints: anti-A/anti-B isoagglutinin titers, rebound, graft function and coordinated transplant timing. [4,8]

Lipopak

Severe familial hypercholesterolemia or progressive cardiovascular disease with elevated Lp(a) despite optimized standard measures, according to local lipoprotein-apheresis eligibility rules.

Endpoints: pre-/post-session LDL-C and Lp(a), inter-session rebound, long-term ASCVD course and vascular-access tolerance. [5,9]

NucleoCore

Aims to adsorb cfDNA/DNA-associated structures implicated in inflammatory and thromboinflammatory disease processes. Mechanistic rationale does not establish routine treatment benefit for sepsis, transplantation, autoimmune disease or oncology.

Endpoints: protocol-defined cfDNA/NET assay and clinical safety outcomes; ethical/regulatory research review where required. [6]

Before every course, record diagnosis, indication strength, failed/available alternatives, patient consent, expected measurable response, a treatment-frequency rationale and discontinuation criteria. Consult disease-specific guidance, local formulary and the actual IFU.

Separated-plasma pathway and regeneration

Conceptual single-column configuration

Patient blood→Blood pump→Plasma separator→Plasma line→POCARD plasma adsorber→Recombination with cells→Patient

The cellular fraction must bypass the plasma adsorber. Use suitable safeguards for air, pressure, leakage, plasma separation quality, temperature and fluid balance. This is a conceptual illustration, not a validated plumbing schematic. [7,8]

Regenerable systems — concept only

Separated plasma→Active adsorption column→Plasma return→Second column / validated regeneration phase (if applicable)
Do not improvise regeneration. Column switching, rinsing, elution, neutralization, preservation, storage, microbiological precautions, number of cycles and final disposal must follow the exact manufacturer and locally approved procedure. Regeneration solutions must never be permitted to enter the patient circuit. Single-use cartridges (ImmunoCore, NucleoCore and brochure-listed 50 mL ABO variants) must not be regenerated. [2–8]

Machine names or conceptual examples mentioned in brochures do not prove that a specific combination is currently approved in the user's country or with the installed console version.

Technical inputs and published manufacturer examples

ParameterWhat the clinician recordsSource-derived boundary / warning
Blood flow (Qb)Set according to access, patient stability and separatorNo universal POCARD plasma-adsorption blood flow demonstrated for all product families.
Plasma flow (Qp)Set by separator extraction fraction, filtration safety and adsorber IFUPOCARD's 2026 transplantation brochure describes ≤60 mL/min for ImmunoCore. This is not a default for other columns. [8]
Processed plasma (L)Target volume routed through adsorberThe brochure gives ABO Adsopak study/protocol examples around 3–3.5 L per cycle, with multi-cycle cases; not a blanket patient prescription. [8]
Treatment timePlanned minutes and independent safety stopping conditionsCalculated from valid Qp and target volume, not from sorbent volume alone.
Session number / frequencyIndication-specific course and laboratory reassessment planManufacturer transplant brochure illustrates a multi-session schedule. Do not extrapolate it to every disease or column. [8]
Regeneration/reuseExact model, patient assignment, cycle counter and IFUImmunoCore and NucleoCore single-use; Immuno-Adsopak, larger ABO Adsopak and Lipopak have model-specific reuse claims. [2–8]
PrimingFull circuit and cartridge dead space, priming solution and disposalExact prime/rinse, orientation and extracorporeal volume: current IFU only.
AnticoagulationDrug, route, dose, lab monitoring and rescue planNo cross-family universal anticoagulant regimen; citrate in impaired metabolism needs particular monitoring.
How to interpret numbers: 60 mL/min is cited for ImmunoCore in one 2026 manufacturer brochure. The 3–3.5 L/cycle example describes an ABO Adsopak transplant pathway. Neither figure is a manufacturer-validated universal prescription, a national regulatory approval, or an automatically safe value for every patient, cartridge size and separator. [8]
Missing specifications must stay missing. Do not guess cartridge pressure limits, maximum extraction fraction, priming anticoagulant, regeneration reagents or treatment duration. Obtain the full current IFU and approved local operating procedure before clinical implementation.

Processed plasma volume & approximate EPV

Calculate the volume of plasma passing through the adsorber. The optionally estimated plasma volume (EPV) is a rough physiologic reference, not a proprietary POCARD treatment dose. These figures do not estimate sorbent capacity or target removal.

Plasma processed—
Plasma flow / hour—
Processed volume / estimated EPV—
Processed plasma (L) = Qp (mL/min) × time (min) ÷ 1,000
Optional EPV approximation: 0.065 × body weight (kg) × (1 − hematocrit fraction) in liters. It is imprecise in marked fluid overload, pregnancy, extreme habitus or unstable patients.
Enter prescribed or otherwise verified inputs to calculate; blank by default to avoid suggesting a standard dose.
Choose a product to view its route and reuse reminder.

Before connection and at initiation

  • Eligibility: confirm disease indication and evidence, exact SKU and local legal authorization, consent, vascular access, current IFU and system/consumable compatibility.
  • Preparation: inspect outer packaging, seal, expiry, storage conditions, patient-assignment record, column orientation and documented priming method. Do not substitute unvalidated solutions.
  • Plasma quality: verify separation performance, pressure limits, and a circuit that prevents cellular blood from entering a plasma-only column.
  • Regenerable column: document patient identity, number of cycles/uses, previous handling, storage/traceability and functional release criteria; never share a regenerated column between patients.
  • Anticoagulation: prescribe per machine/column/center protocol, bleeding risk, coagulation profile and drug contraindications. With regional citrate, monitor ionized calcium and acid–base status; verify whether the system supports it.
  • Medication review: identify therapies potentially removed during adsorption (including Ig therapeutics/IVIG or other protein-bound treatments) and coordinate doses with pharmacy. Check all product-specific drug interaction and warning sections.
  • Circulatory plan: account for full extracorporeal volume, priming/rinse fluid, potential hemodilution, losses and need for any separately ordered replacement strategy.
Do not copy anticoagulation or regeneration regimens from other adsorbers. Publicly available POCARD marketing pages do not provide a complete, cross-compatible operating procedure for these five product families.

Baseline, intraprocedural and post-session checks

WhenCore surveillanceTarget-specific measurements
Pre-treatmentVitals, access check, volume status, CBC/platelets, electrolytes, kidney function, coagulation where relevant; bleeding/infection evaluationIgG/IgM/IgA and relevant antibodies (Ig sorbents); anti-A/anti-B titers (ABO); LDL-C/Lp(a) (Lipopak); research-validated cfDNA/NETs endpoints if used (NucleoCore)
During sessionBlood pressure, symptoms, flow, pressures, clots, air detectors, fluid balance, anticoagulation and calcium if indicatedSelected repeat titers or functional adsorption assessment only when protocol-specified; do not interpret surrogate changes in isolation
End of sessionHemodynamic tolerance, bleeding/hemolysis signs, recovery of circuit blood, relevant lab reassessment and column dispositionPre/post target reduction and possible unintended Ig/protein change, consistent with local testing methodology
Between sessionsResponse, infection risk and hypogammaglobulinemia, renal/hepatic/organ recovery, access patency, treatment harmsAntibody rebound (Ig/ABO), inter-session lipid rebound (Lipopak), protocol-defined endpoints (NucleoCore)

Document frequency of labs and target-response criteria in the patient-specific order. For recurrent Ig adsorption, consider infection susceptibility and Ig depletion. For ABO procedures, coordinate tightly with the transplant center and transfusion laboratory.

Complications, stopping rules and evidence caveats

Procedural risks
  • Hypotension, vasovagal reaction and fluid shifts
  • Access-site bleeding, infection or thrombosis
  • Plasma separator malfunction, clotting or pressure alarms
  • Allergy or biocompatibility reaction
  • Bleeding or electrolyte changes from anticoagulation
  • Potential protein/Ig loss and drug adsorption
Stop or reassess immediately
  • Uncontrolled hypotension, bleeding or severe hypersensitivity
  • Persistent critical pressure, air, leak or clotting alarms
  • Clinically important hemolysis or laboratory deterioration
  • Evidence of regeneration/rinse-fluid crossover
  • Loss of safe vascular access or anticoagulation control
  • Clinical goals no longer justify risk
High-priority verification: A cartridge's apparent adsorption specificity does not eliminate the risk of nonspecific loss of other molecules, hemodynamic effects or catheter-related events. Do not assume a product approved for plasma adsorption in one jurisdiction can be ordered in France/EU without local conformity and hospital authorization.

Evidence maturity: selective anti-A/B adsorption and lipoprotein apheresis have established specialist clinical pathways, but outcomes are context-specific. Immune adsorption efficacy varies by diagnosis. For NucleoCore, the current marketing page documents molecular targeting and possible pathophysiologic settings; that does not itself demonstrate clinical effectiveness or routine indication. [1–9]

Choosing between POCARD plasma sorbents

QuestionImmunoCoreImmuno-AdsopakABO AdsopakLipopakNucleoCore
Primary targetIg / immune complexesIg / immune complexesAnti-A and/or anti-BLDL / Lp(a)cfDNA / DNA structures
Clinical pathwaySelected Ig immunoadsorptionMulti-cycle / series of Ig IAABO-incompatible transplantationLipid apheresisSpecialist/investigational
Column useSingle-usePatient-specific reuseSize-dependentPatient-specific regenerationSingle-use
Monitoring focusAutoantibody / IgAutoantibody / IgAnti-A/B titersLDL-C and Lp(a)Protocol-defined cfDNA assays
Special cautionIg loss, replacement therapy timingRegeneration traceabilityCorrect A/B target and titrationChronic ASCVD/lipid goalsOutcome evidence and protocol approval

This table compares therapeutic logic, not effectiveness; products are not clinically interchangeable. Target-removal performance and safety depend on the protocol, patient, machine, column variant and local IFU.

POCARD separated-plasma adsorption prescription worksheet

Patient / identifier: ______________________________
Date / prescriber: ______________________________
Clinical indication: ______________________________
Evidence and expected endpoint: ________________
Product + exact size/SKU: ______________________
IFU revision / local authorization: _______________
Machine / plasma separator: ____________________
Access / blood flow Qb: ________________________
Plasma flow Qp (mL/min): ______________________
Processed plasma goal (L): _____________________
Time / frequency: ____________________________
Anticoagulation: ______________________________

Safety and completion checklist

For specialist team documentation only. No numerical settings should be signed without the current manufacturer IFU and patient-specific clinical authorization.

Prescription Worksheet — Complete & Print

Fill in this patient-specific prescription worksheet on screen, copy the completed fields, or print the worksheet alone. Blank fields are intentional; the worksheet does not determine appropriate device settings or constitute an authorized medical order.

Prescription fieldClinician entry / verification
Patient & clinical indication
Patient identifier / hospital record
Treatment date / time
Clinical service / prescribing physician
Diagnosis / indication
Therapeutic objective / response criteria
Device & patient assessment
Exact product / modality
Product lot / IFU revision / local authorization
Device platform, separator and extracorporeal circuit
Vascular access
Weight / hematocrit / estimated plasma volume
Treatment prescription
Blood flow and plasma flow
Processed plasma target / treatment volume
Planned duration and treatment frequency
Procedure-specific details
Target-specific marker / baseline and goal
ABO donor / recipient / titers (if applicable)
Column regeneration / reuse protocol
Clinical evidence / protocol approval
Safety, monitoring & authorization
Priming, fluid return and balance
Anticoagulant regimen and monitoring
Medication review / contraindications
Baseline and procedural monitoring
Stopping / escalation criteria
Consent / physician validation / signature

Mandatory clinical verification: the signed institutional order, current model-specific manufacturer IFU, applicable regulatory authorization and patient-specific medical review take precedence. The worksheet remains entirely local to this page; its entries are not saved or transmitted.

Primary links and interpretation of product claims

Source hierarchy: official local market authorization + exact current IFU + approved institutional procedure precede the manufacturer's product webpage, promotional brochure and peer-reviewed reports. All externally linked references require internet access; the guide itself works offline.