Separated-plasma adsorbers only
TheraSorb immunoadsorption separates blood into a cellular fraction and plasma, selectively removes immunoglobulins and other targeted immunological molecules from the plasma in sorbent columns, and returns treated plasma together with blood cells. The columns are not designed for direct whole-blood perfusion. [1,4,5]
Plasma adsorption is not therapeutic plasma exchange (TPE): plasma is normally returned, but some plasma is unavoidably lost to waste during alternating adsorption/regeneration. Do not count processed plasma as exchanged or replaced plasma.
Currently supported prescribing models
TheraSorb® Ig omni 1
Target: IgG1–4, IgA, IgM, IgE, immune complexes, free κ and λ light chains.
Matrix / ligand: 45 mL Sepharose CL-4B with recombinant light-chain-binding proteins.
Configuration: two columns per patient, alternately loaded and regenerated within the session.
Reuse: single-use per treatment; regeneration during the same session does not permit reuse across sessions.
Manufacturer ordering: kit 330-000-964; adsorber reference 330-000-793 (verify item and local catalogue).
2025 leaflet / model IFU. [2,4]
TheraSorb® Ig omni 5
Target: IgG1–4, IgA, IgM, IgE, immune complexes, free κ and λ light chains.
Matrix / ligand: 100 mL Sepharose CL-4B with recombinant light-chain-binding proteins.
Configuration: two columns per patient, alternately loaded and regenerated.
Reuse: up to five treatment sessions for the same patient only, subject to the approved IFU, preservation and contamination checks.
Manufacturer ordering: kit 330-000-965; adsorber reference 330-000-795 (verify item and local catalogue).
2025 leaflet / model IFU. [3,5]
| Attribute | Ig omni 1 | Ig omni 5 |
|---|---|---|
| Therapy | Pan-immunoglobulin / immune-complex adsorption | Pan-immunoglobulin / immune-complex adsorption |
| Plasma-only route | Yes | Yes |
| Pair needed | Two alternating columns | Two alternating columns |
| Sorbent matrix | 45 mL per column | 100 mL per column |
| Between-session reuse | No; single-use | Up to 5 sessions, same patient only, under IFU restrictions |
| Unopened storage | 2025 brochure: room temperature, per label | 2–8 °C, per current label |
| Applicable controller | Miltenyi LIFE and approved disposables | Miltenyi LIFE and approved disposables |
Indications, target molecules and clinical goals
These are non-antigen-specific immunoglobulin adsorbers. They remove broad immunoglobulin classes and related immune complexes; they are not equivalent to antigen-specific ABO-A or ABO-B adsorbers.
Published use includes selected ABO-incompatible transplantation, desensitization, and antibody-mediated rejection. Endpoint should be protocol-specific antibody reduction and graft outcomes, not merely liters processed.
Selected experience includes myasthenia gravis/crisis, severe systemic lupus erythematosus and other antibody-mediated disorders; eligibility and strength of supporting evidence vary greatly by diagnosis.
Manufacturer documentation includes acquired hemophilia with high-titer factor VIII inhibitors among discussed indications. Complex specialist decisions should incorporate concurrent immunosuppression and replacement/hemostatic therapies.
Some manufacturer literature lists indications with weak evidence, including long COVID/ME-CFS and other conditions. Such listings must not be interpreted as proven efficacy, routine indication, or national reimbursement approval.
Prior to approval: record underlying disease, severity and urgency, prior first-line therapy, rationale for adsorption rather than TPE or pharmacologic treatment, planned biomarker/functional endpoints, expected course, immunosuppression implications and stop rules. [2–5]
Plasma separation and alternating columns
Conceptual diagram only: the LIFE platform alternates adsorption and regeneration. Elution, preservation and waste fluids are not returned to the patient. Refer to the actual platform tubing map for its plasma bypass, monitoring transducers, filters, alarms and solution connections. [4,5,10]
Published IFU values versus patient-specific order
Values below are taken from publicly accessible 2022 manufacturer IFUs; they are displayed as verification points, not automatic device settings or recommendations. A newer IFU or platform software can supersede these specifications. [4,5]
| Parameter | Ig omni 1 · published IFU | Ig omni 5 · published IFU | Order requirement |
|---|---|---|---|
| Plasma flow through active column | 5–40 mL/min | 5–40 mL/min | Record approved patient-specific Qp and exact software/IFU limit. |
| Plasma loading per cycle | 50–250 mL | 100–400 mL | Use automatic/model-specific procedure to avoid saturation. |
| Published typical loading | 135 mL/cycle near 30 mL/min | 275 mL/cycle near 30 mL/min | Illustrative IFU examples, not a mandatory clinical set-point. |
| Maximum loading cycles | 60 cycles per session (older IFU) | 60 cycles per session (older IFU) | Check current limits and cumulative plasma loss; do not infer duration from cycles alone. |
| Processed plasma amount | Older IFU discusses 1.5–2.0 patient EPV for certain target depletion goals | Older IFU discusses roughly 1.5–2.0 patient EPV in specific contexts | Document indication, baseline Ig, required response and current IFU before selecting volume. |
| Blood flow / separation ratio | Machine- and separator-dependent | Prescribe Qb, validated extraction fraction, access and separation limit. | |
| Anticoagulation | Local approved approach (e.g., citrate or heparin), with patient-specific contraindications | Specify agent, dosing authority, monitoring, ionized calcium response and stop thresholds. | |
| Storage / preservation | Single-use after session | Same-patient storage at 2–8 °C, specialized preservation and contamination controls | Confirm exact IFU, traceability, storage records and eligibility before reuse. |
Processed plasma and estimated patient plasma volume
For documentation only. Enter the average plasma flow during active adsorption and the cumulative active adsorption time. Total appointment or machine running time will overestimate processing if it includes regeneration pauses, standby or bypass.
Patient plasma volume estimate (L) ≈ 0.065 × weight (kg) × [1 − hematocrit / 100]
Ig omni 1: single-use column pair; regeneration within the same session only.
Enter the values; no clinical setting is recommended by this tool.
Access, equipment, priming and medications
Before connecting the patient
- Confirm diagnosis, consent, treatment goals and expert responsibility.
- Identify model, pair, batch, expiration, IFU revision, approved LIFE machine/software and compatible tubing sets.
- For Ig omni 5, confirm same-patient identity, prior-use count, storage temperature, preservation records and microbiological/column integrity checks.
- Check extracorporeal volume, vascular access suitability, baseline blood count, immunoglobulins, coagulation and clinical status.
- Follow automated device integrity testing, separation/air/particle filter and model-specific solutions management.
Medication and infection review
- Check ACE inhibitors and medications affecting bradykinin; older IFUs caution about enhanced hypotension with ACE-inhibitor medication. Do not discontinue drugs without prescriber authorization.
- Plan timing of IVIg, therapeutic monoclonal antibodies and other biologics; adsorption can remove therapeutic antibodies.
- Screen for severe immunodeficiency, systemic infection and patient-specific infection vulnerability.
- Review potential allergic responses to matrix/ligands, bleeding risk, anticoagulation restrictions and severe hemodynamic instability.
Important safety precautions
- Active generalized infection or severe immune deficiency (per specific IFU)
- Inability to tolerate extracorporeal volume / plasma component losses
- Hypercoagulability or inability to deliver acceptable anticoagulation
- Suspected ligand/matrix hypersensitivity
- Hemodynamic instability and impaired vascular access
- Allergic, anaphylactoid and bradykinin-related reactions; hypotension
- Symptomatic hypocalcemia, dysrhythmia or acid–base changes if citrate is used
- Thrombocytopenia, bleeding, access complications, hemolysis
- Ig depletion with increased infection risk or loss of protective antibodies
- Albumin/protein loss, clotting or adsorber-pressure alarms
The published IFUs include model-specific absolute or relative contraindications. Recheck them directly — this list is an overview, not the complete label. [4,5]
Before, during and after each session
| Phase | Clinical and laboratory review | Document |
|---|---|---|
| Baseline | Vital signs, volume status, access, CBC/platelets, disease-specific antibody panel, IgG/IgM/IgA as appropriate, coagulation/fibrinogen, calcium and other labs indicated by clinical context. | Clinical baseline, target, consent, contraindication/medication review. |
| During adsorption | Continuous hemodynamics and symptoms, circuit blood/plasma flows, separation pressures, adsorber differential pressure, alarms and clotting signs; anticoagulation and ionized calcium when relevant. | Active adsorption duration, column switches, cycles, fluid/plasma loss, adverse events. |
| End of session | Tolerance, bleeding/access sites, selected labs for Ig/antibodies, hematology, protein and calcium status. Recheck drugs/biologics that may have been adsorbed. | Actual processed volume, response, follow-up plan and safe reinfusion/disconnection. |
| Across sessions | Clinical endpoints, antibody rebound, immunosuppression, infection risk and need for alternative/adjunctive therapy. | Decision to continue, stop, change plan; for Ig omni 5 verify remaining permitted use count. |
When to pause, stop or urgently reassess
- Significant hypotension, shock, chest pain, severe arrhythmia or respiratory symptoms
- Suspected severe hypersensitivity or other acute systemic reaction
- Severe symptomatic hypocalcemia, active bleeding or possible hemolysis
- Air, leakage, clotting, pressure or separation alarms that cannot be safely resolved
- Any concern that elution or preservation fluid could enter the patient line
- Target clinical response achieved or an alternative therapy becomes preferable
- Disproportionate immunoglobulin depletion or infection risk
- Insufficient response after agreed assessment period
- Unacceptable cumulative plasma/protein losses
- Prohibited or unverified Ig omni 5 reuse condition
Follow the LIFE alarm and emergency instructions and local response pathway. An education guide cannot replace the manufacturer-defined stop conditions.
Ig omni 1 versus Ig omni 5 versus TPE
| Feature | Ig omni 1 | Ig omni 5 | TPE (comparison only) |
|---|---|---|---|
| Therapy principle | Pan-Ig / light-chain adsorption | Pan-Ig / light-chain adsorption | Non-selective plasma removal |
| Separated plasma processed | Yes | Yes | Yes |
| Adsorber | Two columns; 45 mL sorbent each | Two columns; 100 mL sorbent each | No adsorption column required |
| Full replacement-fluid exchange | No, though some plasma is lost during regeneration | No, though some plasma is lost during regeneration | Yes |
| Reuse | Single treatment | Up to 5 same-patient sessions under IFU | Not applicable |
This table is a technical distinction, not a claim that one modality is superior or interchangeable for a given disease.
Printable prescription checklist
For clinical documentation support only; all treatment settings require current IFU and specialist authorization.
Prescription Worksheet — Complete & Print
Fill in this patient-specific prescription worksheet on screen, copy the completed fields, or print the worksheet alone. Blank fields are intentional; the worksheet does not determine appropriate device settings or constitute an authorized medical order.
Mandatory clinical verification: the signed institutional order, current model-specific manufacturer IFU, applicable regulatory authorization and patient-specific medical review take precedence. The worksheet remains entirely local to this page; its entries are not saved or transmitted.
Manufacturer documents and supporting evidence
- Miltenyi Biotec — current TheraSorb Therapeutic Apheresis portfolio. Current manufacturer overview of the plasma immunoadsorption line and LIFE platform; references Ig omni 1 and 5.
- Miltenyi Biotec — TheraSorb Ig omni 1 product brochure (2025). Single-use adsorber, 45 mL Sepharose, Ig/light-chain targets; kit 330-000-964.
- Miltenyi Biotec — TheraSorb Ig omni 5 product brochure (2025). Five-treatment device, 100 mL Sepharose, Ig/light-chain targets; kit 330-000-965.
- Miltenyi Biotec — TheraSorb Ig omni 1 IFU (public copy, 2022). Single-use pair, plasma loading/flow, maximum cycles, contraindications and precautions. Current local revision takes precedence.
- Miltenyi Biotec — TheraSorb Ig omni 5 IFU (public copy, 2022). Same-patient five-session use, regeneration/preservation, fluid and pressure constraints, medicine interactions.
- Miltenyi Bioindustry — regulatory notes, TheraSorb product family. EU CE marking and country-dependent US status; not confirmation of a specific hospital's procurement.
- Miltenyi Biotec — LIFE 21 change notification CC-21-0447. Historical documentation of Ig omni 1/5, LDL, LDL pro and IgE models; not a verified 2026 price list.
- Röseler et al. (2025), Journal of Clinical Apheresis — combined lipoprotein apheresis/hemodialysis. Documents clinical use of TheraSorb LDL; not proof of a current manufacturer order listing.
- IgE immunoadsorption: technical background and clinical experience (2024). Describes historical TheraSorb IgE clinical methodology.
- Miltenyi Biotec — Immunoadsorption product area. Manufacturer's current plasma immunoadsorption category.
- Miltenyi Biotec — LIFE 21 general user manual. Technical operating framework; always use the current locally supplied version.