Clinical apheresis · Practical reference

Miltenyi Biotec
Plasma Adsorption Prescription Guide

TheraSorb® Ig omni 1 · TheraSorb® Ig omni 5
Manufacturer-documented immunoadsorption of separated human plasma. Product selection, circuit safety, evidence boundaries, plasma-volume calculator and printable prescription checklist.

Clinical use limitation. Independent educational framework only — not a Miltenyi Biotec instruction for use (IFU), treatment authorization, or executable medical order. Confirm the exact model, current country-specific IFU, software version and compatibility of the LIFE apheresis platform before prescribing. Manufacturer IFUs take priority for contraindications, compatible kits, priming, flow, anticoagulation, cycle limits, pressure limits, reuse, and storage. All treatments require specialist judgment and institutional authorization. [1–6]
Current-market screening: Ig omni 1 and Ig omni 5 have manufacturer product leaflets dated 2025 and current manufacturer TheraSorb portfolio references. Their current ordering status, supply and authorization in a particular country still require written confirmation from Miltenyi Biotec. [1–3,6]

Separated-plasma adsorbers only

TheraSorb immunoadsorption separates blood into a cellular fraction and plasma, selectively removes immunoglobulins and other targeted immunological molecules from the plasma in sorbent columns, and returns treated plasma together with blood cells. The columns are not designed for direct whole-blood perfusion. [1,4,5]

Plasma separationImmunoadsorption columnLIFE platformIg omni 1 / Ig omni 5No whole-blood cartridges

Plasma adsorption is not therapeutic plasma exchange (TPE): plasma is normally returned, but some plasma is unavoidably lost to waste during alternating adsorption/regeneration. Do not count processed plasma as exchanged or replaced plasma.

Historical / unconfirmed-for-current-ordering models: TheraSorb LDL, LDL pro, and TheraSorb IgE are documented as plasma sorbents in older Miltenyi technical notifications, clinical studies and device records, but current manufacturer ordering information for these exact models could not be verified as of this review. They are therefore not included in the active model prescriptions. The predecessor Ig flex / Ig pro names are likewise not substituted for the 2025 Ig omni products. This is not proof of discontinuation; confirm directly with the manufacturer if considering procurement. [1,7–9]

Currently supported prescribing models

Single-use column pair

TheraSorb® Ig omni 1

Target: IgG1–4, IgA, IgM, IgE, immune complexes, free κ and λ light chains.

Matrix / ligand: 45 mL Sepharose CL-4B with recombinant light-chain-binding proteins.

Configuration: two columns per patient, alternately loaded and regenerated within the session.

Reuse: single-use per treatment; regeneration during the same session does not permit reuse across sessions.

Manufacturer ordering: kit 330-000-964; adsorber reference 330-000-793 (verify item and local catalogue).

2025 leaflet / model IFU. [2,4]

Reusable, single-patient column pair

TheraSorb® Ig omni 5

Target: IgG1–4, IgA, IgM, IgE, immune complexes, free κ and λ light chains.

Matrix / ligand: 100 mL Sepharose CL-4B with recombinant light-chain-binding proteins.

Configuration: two columns per patient, alternately loaded and regenerated.

Reuse: up to five treatment sessions for the same patient only, subject to the approved IFU, preservation and contamination checks.

Manufacturer ordering: kit 330-000-965; adsorber reference 330-000-795 (verify item and local catalogue).

2025 leaflet / model IFU. [3,5]

AttributeIg omni 1Ig omni 5
TherapyPan-immunoglobulin / immune-complex adsorptionPan-immunoglobulin / immune-complex adsorption
Plasma-only routeYesYes
Pair neededTwo alternating columnsTwo alternating columns
Sorbent matrix45 mL per column100 mL per column
Between-session reuseNo; single-useUp to 5 sessions, same patient only, under IFU restrictions
Unopened storage2025 brochure: room temperature, per label2–8 °C, per current label
Applicable controllerMiltenyi LIFE and approved disposablesMiltenyi LIFE and approved disposables
Important terminology: “Single-use” does not mean “single-pass.” Ig omni 1 and Ig omni 5 use repeated adsorption/regeneration cycles within a treatment; only Ig omni 5 is intended for controlled between-session reuse in the same patient. [4,5]

Indications, target molecules and clinical goals

These are non-antigen-specific immunoglobulin adsorbers. They remove broad immunoglobulin classes and related immune complexes; they are not equivalent to antigen-specific ABO-A or ABO-B adsorbers.

Antibody-mediated renal/transplant conditions

Published use includes selected ABO-incompatible transplantation, desensitization, and antibody-mediated rejection. Endpoint should be protocol-specific antibody reduction and graft outcomes, not merely liters processed.

Autoimmune neurologic and systemic diseases

Selected experience includes myasthenia gravis/crisis, severe systemic lupus erythematosus and other antibody-mediated disorders; eligibility and strength of supporting evidence vary greatly by diagnosis.

Inhibitor-associated hematology / other selected uses

Manufacturer documentation includes acquired hemophilia with high-titer factor VIII inhibitors among discussed indications. Complex specialist decisions should incorporate concurrent immunosuppression and replacement/hemostatic therapies.

Investigational or low-evidence contexts

Some manufacturer literature lists indications with weak evidence, including long COVID/ME-CFS and other conditions. Such listings must not be interpreted as proven efficacy, routine indication, or national reimbursement approval.

Prior to approval: record underlying disease, severity and urgency, prior first-line therapy, rationale for adsorption rather than TPE or pharmacologic treatment, planned biomarker/functional endpoints, expected course, immunosuppression implications and stop rules. [2–5]

Plasma separation and alternating columns

Patient blood→Blood pump→Plasma separation→Plasma pump→Active TheraSorb column→Plasma + cells recombined→Patient return
Standby column→Plasma displacement→Elution / buffer neutralization→Waste and regeneration→Ready for switch

Conceptual diagram only: the LIFE platform alternates adsorption and regeneration. Elution, preservation and waste fluids are not returned to the patient. Refer to the actual platform tubing map for its plasma bypass, monitoring transducers, filters, alarms and solution connections. [4,5,10]

Critical contamination prevention: Ig omni 5 IFU warns that the preservation buffer containing azide is toxic if inadvertently infused. The manufacturer instructs that preservation solution be handled after the patient is disconnected and fully removed through the IFU-specified preparation/rinsing steps before subsequent use. Never improvise buffer preparation or line routing. [5]

Published IFU values versus patient-specific order

Values below are taken from publicly accessible 2022 manufacturer IFUs; they are displayed as verification points, not automatic device settings or recommendations. A newer IFU or platform software can supersede these specifications. [4,5]

ParameterIg omni 1 · published IFUIg omni 5 · published IFUOrder requirement
Plasma flow through active column5–40 mL/min5–40 mL/minRecord approved patient-specific Qp and exact software/IFU limit.
Plasma loading per cycle50–250 mL100–400 mLUse automatic/model-specific procedure to avoid saturation.
Published typical loading135 mL/cycle near 30 mL/min275 mL/cycle near 30 mL/minIllustrative IFU examples, not a mandatory clinical set-point.
Maximum loading cycles60 cycles per session (older IFU)60 cycles per session (older IFU)Check current limits and cumulative plasma loss; do not infer duration from cycles alone.
Processed plasma amountOlder IFU discusses 1.5–2.0 patient EPV for certain target depletion goalsOlder IFU discusses roughly 1.5–2.0 patient EPV in specific contextsDocument indication, baseline Ig, required response and current IFU before selecting volume.
Blood flow / separation ratioMachine- and separator-dependentPrescribe Qb, validated extraction fraction, access and separation limit.
AnticoagulationLocal approved approach (e.g., citrate or heparin), with patient-specific contraindicationsSpecify agent, dosing authority, monitoring, ionized calcium response and stop thresholds.
Storage / preservationSingle-use after sessionSame-patient storage at 2–8 °C, specialized preservation and contamination controlsConfirm exact IFU, traceability, storage records and eligibility before reuse.
Plasma loss matters: both alternation/regeneration processes divert a nonzero amount of plasma to waste. The gross amount pumped through the adsorbers is not a replacement volume and may differ from the plasma effectively processed. Monitor patient fluid/protein losses rather than assuming complete conservation. [4,5]

Processed plasma and estimated patient plasma volume

For documentation only. Enter the average plasma flow during active adsorption and the cumulative active adsorption time. Total appointment or machine running time will overestimate processing if it includes regeneration pauses, standby or bypass.

Estimated processed plasma—
Estimated patient plasma volume—
Processed / patient plasma volume—
Processed plasma (L) = Qp (mL/min) × active time (min) ÷ 1,000
Patient plasma volume estimate (L) ≈ 0.065 × weight (kg) × [1 − hematocrit / 100]

Ig omni 1: single-use column pair; regeneration within the same session only.

Enter the values; no clinical setting is recommended by this tool.

Access, equipment, priming and medications

Before connecting the patient

  • Confirm diagnosis, consent, treatment goals and expert responsibility.
  • Identify model, pair, batch, expiration, IFU revision, approved LIFE machine/software and compatible tubing sets.
  • For Ig omni 5, confirm same-patient identity, prior-use count, storage temperature, preservation records and microbiological/column integrity checks.
  • Check extracorporeal volume, vascular access suitability, baseline blood count, immunoglobulins, coagulation and clinical status.
  • Follow automated device integrity testing, separation/air/particle filter and model-specific solutions management.

Medication and infection review

  • Check ACE inhibitors and medications affecting bradykinin; older IFUs caution about enhanced hypotension with ACE-inhibitor medication. Do not discontinue drugs without prescriber authorization.
  • Plan timing of IVIg, therapeutic monoclonal antibodies and other biologics; adsorption can remove therapeutic antibodies.
  • Screen for severe immunodeficiency, systemic infection and patient-specific infection vulnerability.
  • Review potential allergic responses to matrix/ligands, bleeding risk, anticoagulation restrictions and severe hemodynamic instability.
Do not copy a priming recipe from this guide. The manufacturer IFU specifies different flush volumes, buffer compositions and flow steps by model and reuse status. Wrong line placement or incomplete rinsing can cause a serious patient safety event. Follow the exact device instructions and competent-team checklist. [4,5]

Important safety precautions

Patient / disease factors
  • Active generalized infection or severe immune deficiency (per specific IFU)
  • Inability to tolerate extracorporeal volume / plasma component losses
  • Hypercoagulability or inability to deliver acceptable anticoagulation
  • Suspected ligand/matrix hypersensitivity
  • Hemodynamic instability and impaired vascular access
Procedure-specific hazards
  • Allergic, anaphylactoid and bradykinin-related reactions; hypotension
  • Symptomatic hypocalcemia, dysrhythmia or acid–base changes if citrate is used
  • Thrombocytopenia, bleeding, access complications, hemolysis
  • Ig depletion with increased infection risk or loss of protective antibodies
  • Albumin/protein loss, clotting or adsorber-pressure alarms

The published IFUs include model-specific absolute or relative contraindications. Recheck them directly — this list is an overview, not the complete label. [4,5]

Before, during and after each session

PhaseClinical and laboratory reviewDocument
BaselineVital signs, volume status, access, CBC/platelets, disease-specific antibody panel, IgG/IgM/IgA as appropriate, coagulation/fibrinogen, calcium and other labs indicated by clinical context.Clinical baseline, target, consent, contraindication/medication review.
During adsorptionContinuous hemodynamics and symptoms, circuit blood/plasma flows, separation pressures, adsorber differential pressure, alarms and clotting signs; anticoagulation and ionized calcium when relevant.Active adsorption duration, column switches, cycles, fluid/plasma loss, adverse events.
End of sessionTolerance, bleeding/access sites, selected labs for Ig/antibodies, hematology, protein and calcium status. Recheck drugs/biologics that may have been adsorbed.Actual processed volume, response, follow-up plan and safe reinfusion/disconnection.
Across sessionsClinical endpoints, antibody rebound, immunosuppression, infection risk and need for alternative/adjunctive therapy.Decision to continue, stop, change plan; for Ig omni 5 verify remaining permitted use count.

When to pause, stop or urgently reassess

Immediate safety triggers
  • Significant hypotension, shock, chest pain, severe arrhythmia or respiratory symptoms
  • Suspected severe hypersensitivity or other acute systemic reaction
  • Severe symptomatic hypocalcemia, active bleeding or possible hemolysis
  • Air, leakage, clotting, pressure or separation alarms that cannot be safely resolved
  • Any concern that elution or preservation fluid could enter the patient line
Course-level stopping criteria
  • Target clinical response achieved or an alternative therapy becomes preferable
  • Disproportionate immunoglobulin depletion or infection risk
  • Insufficient response after agreed assessment period
  • Unacceptable cumulative plasma/protein losses
  • Prohibited or unverified Ig omni 5 reuse condition

Follow the LIFE alarm and emergency instructions and local response pathway. An education guide cannot replace the manufacturer-defined stop conditions.

Ig omni 1 versus Ig omni 5 versus TPE

FeatureIg omni 1Ig omni 5TPE (comparison only)
Therapy principlePan-Ig / light-chain adsorptionPan-Ig / light-chain adsorptionNon-selective plasma removal
Separated plasma processedYesYesYes
AdsorberTwo columns; 45 mL sorbent eachTwo columns; 100 mL sorbent eachNo adsorption column required
Full replacement-fluid exchangeNo, though some plasma is lost during regenerationNo, though some plasma is lost during regenerationYes
ReuseSingle treatmentUp to 5 same-patient sessions under IFUNot applicable

This table is a technical distinction, not a claim that one modality is superior or interchangeable for a given disease.

Printable prescription checklist

For clinical documentation support only; all treatment settings require current IFU and specialist authorization.

Prescription Worksheet — Complete & Print

Fill in this patient-specific prescription worksheet on screen, copy the completed fields, or print the worksheet alone. Blank fields are intentional; the worksheet does not determine appropriate device settings or constitute an authorized medical order.

Prescription fieldClinician entry / verification
Patient & clinical indication
Patient identifier / hospital record
Treatment date / time
Clinical service / prescribing physician
Diagnosis / indication
Therapeutic objective / response criteria
Device & patient assessment
Exact product / modality
Product lot / IFU revision / local authorization
Device platform, separator and extracorporeal circuit
Vascular access
Weight / hematocrit / estimated plasma volume
Treatment prescription
Blood flow and plasma flow
Processed plasma target / treatment volume
Planned duration and treatment frequency
Procedure-specific details
Target Ig / antibody and baseline level
LIFE system and column configuration
Cycle program and column reuse count
ACE-inhibitor / bradykinin and drug review
Safety, monitoring & authorization
Priming, fluid return and balance
Anticoagulant regimen and monitoring
Medication review / contraindications
Baseline and procedural monitoring
Stopping / escalation criteria
Consent / physician validation / signature

Mandatory clinical verification: the signed institutional order, current model-specific manufacturer IFU, applicable regulatory authorization and patient-specific medical review take precedence. The worksheet remains entirely local to this page; its entries are not saved or transmitted.

Manufacturer documents and supporting evidence

  1. Miltenyi Biotec — current TheraSorb Therapeutic Apheresis portfolio. Current manufacturer overview of the plasma immunoadsorption line and LIFE platform; references Ig omni 1 and 5.
  2. Miltenyi Biotec — TheraSorb Ig omni 1 product brochure (2025). Single-use adsorber, 45 mL Sepharose, Ig/light-chain targets; kit 330-000-964.
  3. Miltenyi Biotec — TheraSorb Ig omni 5 product brochure (2025). Five-treatment device, 100 mL Sepharose, Ig/light-chain targets; kit 330-000-965.
  4. Miltenyi Biotec — TheraSorb Ig omni 1 IFU (public copy, 2022). Single-use pair, plasma loading/flow, maximum cycles, contraindications and precautions. Current local revision takes precedence.
  5. Miltenyi Biotec — TheraSorb Ig omni 5 IFU (public copy, 2022). Same-patient five-session use, regeneration/preservation, fluid and pressure constraints, medicine interactions.
  6. Miltenyi Bioindustry — regulatory notes, TheraSorb product family. EU CE marking and country-dependent US status; not confirmation of a specific hospital's procurement.
  7. Miltenyi Biotec — LIFE 21 change notification CC-21-0447. Historical documentation of Ig omni 1/5, LDL, LDL pro and IgE models; not a verified 2026 price list.
  8. Röseler et al. (2025), Journal of Clinical Apheresis — combined lipoprotein apheresis/hemodialysis. Documents clinical use of TheraSorb LDL; not proof of a current manufacturer order listing.
  9. IgE immunoadsorption: technical background and clinical experience (2024). Describes historical TheraSorb IgE clinical methodology.
  10. Miltenyi Biotec — Immunoadsorption product area. Manufacturer's current plasma immunoadsorption category.
  11. Miltenyi Biotec — LIFE 21 general user manual. Technical operating framework; always use the current locally supplied version.
Source hierarchy: current IFU + local device approval and operator training first; then platform software instructions; then product marketing and peer-reviewed evidence. If any current supplier listing conflicts with this review, update the included model list before prescribing. References need internet access; the HTML itself functions offline.