Clinical apheresis · Practical reference

Jafron Plasma Adsorption
Prescription Guide

Plasma-route adsorption only · BS-series bilirubin columns · BS330 + HA330-II (DPMAS) · DNA230 plasma immunoadsorption
Clinical evidence, circuit planning, surveillance and a printable prescription checklist · November 2026

Clinical-use limitation. This is an educational prescribing framework—not a manufacturer IFU, validated treatment order, or product approval statement. This guide covers plasma-route configurations only. Use a cartridge in a separated-plasma circuit only if its exact model and market-specific IFU permit that configuration. Indications, availability, priming, flow/pressure limits, anticoagulation, reuse prohibition and alarm thresholds must be confirmed locally. Do not transfer settings between Jafron models.

Scope and treatment objectives

Plasma adsorption (PA) separates plasma from cellular blood components and passes the separated plasma through an adsorbent before reinfusion. Unlike therapeutic plasma exchange (TPE), the processed plasma is generally returned rather than discarded and fully replaced.

  • BS-series: resin-based adsorption of bilirubin and bile acids in selected severe hyperbilirubinemia/liver failure pathways.
  • DPMAS: double plasma molecular adsorption using BS330 plus HA330-II, usually connected in series in the plasma branch, designed to target bilirubin/bile acids and additional adsorbable substances.
  • DNA230: DNA-based immunoadsorbent targeting certain anti-DNA/antinuclear antibodies and immune complexes; use and supporting evidence are indication- and market-dependent.
Evidence interpretation: Much of the DPMAS literature consists of observational cohorts, pilot studies, and region-specific expert guidance. Biochemical bilirubin reduction does not by itself establish a survival benefit. These therapies require multidisciplinary case selection, especially in acute-on-chronic liver failure.

Jafron adsorbers with a documented plasma-circuit configuration

Product / familyAdsorption targetPlasma-route relevanceCritical verification
BS330 (BS series)Bilirubin, bile acids; anion-exchange adsorbentDedicated plasma bilirubin adsorption column; can be used in validated PA/DPMAS circuitsExact device IFU, compatibility with separator and pump, approved indications
HA330-II (in DPMAS)Broad resin adsorption of additional toxins and inflammatory mediators in selected liver-support protocolsDocumented in a separated-plasma pathway when used with BS330 as DPMASVerify that the exact device, separator and BS330 + HA330-II circuit are permitted locally; combination settings must be validated
DNA230Anti-dsDNA/ANA-associated autoantibodies and immune complexesManufacturer brochure illustrates a separated-plasma immunoadsorption pathway with DNA230Confirm indication, plasma-route authorization, compatibility, evidence limitations and IFU
BS80 / BS380Bilirubin adsorption, model-specificBoth are listed as plasma bilirubin adsorption columns; specifications differ by modelConfirm exact model IFU, local approval, clinical availability and compatibility before use
Included products are limited to documented separated-plasma adsorption configurations. Regulatory status, marketed models and authorized circuits may differ by country and IFU revision. [1–3]

Patient selection and treatment goals

BS330 / DPMAS

Severe hyperbilirubinemia or bile acid accumulation within a specialist liver support pathway; selected acute liver injury or acute-on-chronic liver failure situations.

Goals: bilirubin/bile acid reduction, symptom improvement, organ support and bridging decisions.

DNA230

Selected refractory autoimmune disease pathways described in manufacturer literature, especially systemic lupus erythematosus; evidence and regulatory indication require case-by-case verification.

Goals: selective removal of target antibodies/immune complexes and disease-specific response assessment.

Before ordering: document diagnosis, clinical severity, conventional treatment already provided, expected therapeutic endpoint, alternatives (including TPE), patient consent, and a predefined reassessment/stop plan. Assess transplant eligibility promptly when relevant.

Technical configuration

Typical DPMAS conceptual circuit

Patient blood→Blood pump→Plasma separator→Plasma pump→BS330→HA330-II→Plasma recombination→Patient return

Important: the two adsorption columns in a commonly published DPMAS setup are in series, not parallel. The cellular fraction bypasses the plasma adsorbers and rejoins the treated plasma prior to return. Actual circuit routing, air traps, safety detectors and pressure transducers depend on the machine and IFUs.

Single-column PA

Blood circuit→Plasma separation→BS-series plasma adsorber→Plasma reinfusion

Document whether replacement fluid is planned. Unlike TPE, isolated plasma adsorption ordinarily returns processed plasma rather than replacing the full treated volume; fluid/protein losses can still require clinical management.

Prescription inputs and published examples

ParameterIllustrative published DPMAS valuesPrescription rule
Blood flow100–150 mL/min in a 2024 expert technical specification; 100–130 mL/min in one trial protocolSpecify per device/vascular access and patient tolerance
Plasma flow20–45 mL/min in expert guidance; 1,200–1,500 mL/h in a published trial protocolSet according to separator extraction fraction, hemoconcentration and pressure limits
Plasma separation fractionApproximately 20–30% in cited membrane techniquesValidate against hematocrit and separator IFU; not a universal set-point
Plasma processed per sessionApproximately 5–6 L in Chinese expert guidance; study protocols varyNot equivalent to plasma volume removed or replaced
Session durationOften 2–4 hours in published DPMAS methodsUse prescriber order, clinical response and device capacity/limits
Number/frequencyVariable: expert guidance reports multiple sessions with frequent reassessmentNo universal course; avoid indiscriminate daily treatment
Replacement fluidNo routine full-volume plasma replacement in isolated adsorptionDocument volume balance, albumin changes and whether sequential PE is separately ordered
Source/approval boundary: All numbers above are examples reported in specific studies or expert guidance, not default Jafron machine settings or manufacturer-certified limits. The available manufacturer literature does not establish one universal flow, dose, anticoagulation or treatment duration for every Jafron adsorption cartridge. [4–6]

Processed plasma volume estimator

Estimates the quantity of plasma circulating through the adsorber; it is not the amount of plasma discarded, nor a validated Jafron dose calculator.

Processed plasma5.40 L
Rate per hour1.80 L/h
Per kg body weight77 mL/kg
Processed plasma (L) = Plasma flow (mL/min) × Duration (min) ÷ 1,000

Estimates are illustrative and must not be used to prescribe without clinical verification.

Device setup, priming and anticoagulation

  • Confirm the exact cartridge code, batch/expiry, package integrity, single-use status, separator compatibility, and locally authorized indication.
  • Apply manufacturer-specific installation and priming steps; do not reproduce experimental heparin-soaking recipes as standard clinical instructions.
  • Use strict aseptic connection technique; check adsorber orientation, degassing, bypass/recombination, air detection, pressure lines, and circuit integrity before connection.
  • Choose anticoagulation (e.g., heparin, regional citrate, or another institutional strategy) based on the circuit, liver function, bleeding risk, HIT history, and approved protocol.
  • When citrate is used, monitor ionized calcium and acid-base status; in liver dysfunction, ensure appropriate expertise and enhanced monitoring.
  • Document circuit blood volume, priming/displacement fluid, and fluid balance, particularly in children, anemia, and hemodynamic instability.
No generic anticoagulant prescription: published studies use different heparin or other agents, and the F20 device manual explicitly identifies supported anticoagulation for its configurations. A physician-approved protocol and the applicable equipment IFU take precedence. [3,5]

Pre-treatment, intraprocedural and follow-up assessment

TimeAssessment
Before each sessionClinical examination, vital signs, weight and fluid status; CBC/platelets; bilirubin (total/direct), liver function, renal function, electrolytes, coagulation and fibrinogen as indicated; calcium when relevant; vascular access and medication review
During treatmentBlood pressure, heart rate, consciousness, symptoms; blood/plasma flow; separator and adsorber pressures, transmembrane pressure, clotting signs, air alarms and fluid balance; anticoagulation monitoring as appropriate
At session endClinical tolerance, bleeding/hemolysis signs, residual circuit blood management; selected repeat labs, especially bilirubin, blood count and coagulation depending on patient and technique
Across treatment courseTrajectory of bilirubin/bile acids, organ function and symptoms; thrombocytopenia/protein loss; objective response relative to prespecified goals; decision to continue, switch modality, or stop

Adsorption may influence drugs or protein-bound substances; review therapeutic drug monitoring and the timing of essential medications with clinical pharmacy when relevant.

Complications, contraindication checks and escalation

Key procedural risks
  • Hypotension or volume shifts
  • Bleeding, thrombocytopenia or coagulation abnormalities
  • Clotting, increased pressure or filter failure
  • Access infection, bleeding or thrombosis
  • Allergic/biocompatibility reactions
  • Hypocalcemia or acid–base changes with citrate
Stop/reassess immediately when
  • Severe hemodynamic instability develops
  • Uncontrolled bleeding or suspected severe reaction occurs
  • Persistent air, pressure or clotting alarm cannot be safely resolved
  • Clinically significant lab deterioration occurs
  • Therapy no longer meets goals or risks outweigh benefit

Product contraindications and warnings are model-specific and must be checked against the applicable IFU. Screening for severe active bleeding, DIC, hypersensitivity, and feasibility of anticoagulation is particularly important.

PA vs DPMAS vs TPE: key differences

ModeWhat is processedPlasma replacementTypical purpose
BS-series single-column PASeparated plasma through a bilirubin adsorberNot generally a full-plasma-volume replacementTargeted bilirubin/bile acid adsorption
DPMASSeparated plasma through BS330 and HA330-II in seriesNot routine in adsorption phase; may be combined sequentially with PEMultiple adsorbable targets in selected liver support protocols
DNA230 plasma immunoadsorptionSeparated plasma through a validated immunoadsorbent configurationUsually returned after adsorption if configured as PASelected antibody/immune-complex removal
TPEPatient plasma removedYes, albumin and/or plasmaBroad removal or replacement of plasma components

Do not count liters of processed plasma as liters of exchanged plasma. Do not assume the same treatment goals or evidence base apply across these modes.

Prescription checklist

For use as a documentation aid only; patient-specific orders require qualified clinical review.

Prescription Worksheet — Complete & Print

Fill in this patient-specific prescription worksheet on screen, copy the completed fields, or print the worksheet alone. Blank fields are intentional; the worksheet does not determine appropriate device settings or constitute an authorized medical order.

Prescription fieldClinician entry / verification
Patient & clinical indication
Patient identifier / hospital record
Treatment date / time
Clinical service / prescribing physician
Diagnosis / indication
Therapeutic objective / response criteria
Device & patient assessment
Exact product / modality
Product lot / IFU revision / local authorization
Device platform, separator and extracorporeal circuit
Vascular access
Weight / hematocrit / estimated plasma volume
Treatment prescription
Blood flow and plasma flow
Processed plasma target / treatment volume
Planned duration and treatment frequency
Procedure-specific details
Target: bilirubin, bile acids or antibodies
BS330 + HA330-II serial DPMAS route (if used)
Cartridge combination / device compatibility
Organ support / escalation plan
Safety, monitoring & authorization
Priming, fluid return and balance
Anticoagulant regimen and monitoring
Medication review / contraindications
Baseline and procedural monitoring
Stopping / escalation criteria
Consent / physician validation / signature

Mandatory clinical verification: the signed institutional order, current model-specific manufacturer IFU, applicable regulatory authorization and patient-specific medical review take precedence. The worksheet remains entirely local to this page; its entries are not saved or transmitted.

Selected references and source limitations

  1. Jafron — Therapies & Products. Official manufacturer overview; consult the exact product IFU.
  2. Jafron plasma adsorption product brochure. Manufacturer illustrations of the BS330 plasma route, BS330 + HA330-II DPMAS, and DNA230 plasma immunoadsorption; not a substitute for IFU.
  3. BS330 Plasma Bilirubin Adsorption Column information. Manufacturer specifications; warnings and contraindications require the product IFU.
  4. Clinical application and technical operation specification of DPMAS. Expert technical guidance (2024); primarily Chinese-language.
  5. Nafamostat anticoagulation in DPMAS plus sequential half-dose plasma exchange. Trial protocol (2025); study-specific settings, not a universal clinical prescription.
  6. Association between processed plasma volume and bilirubin reduction. Clinical observational study of BS330 + HA330-II DPMAS (2025).
  7. Jafron DNA230 immunoadsorbent column. Manufacturer description; plasma-route compatibility and indications require local confirmation.
  8. Integration of a blood return strategy into DPMAS. Published separated-plasma circuit description (2023).
Evidence hierarchy: prioritize local regulatory authorization, current cartridge and machine IFUs and institution-approved protocols first; then disease-specific guidance and peer-reviewed trials; use product brochures only to identify products and conceptual modes.