Scope and mechanism of plasma adsorption
This guide covers adsorber cartridges through which separated plasma flows. Blood cells are separated upstream, the plasma passes through a sorbent that selectively binds target substances, then treated plasma is recombined with cells and returned to the patient. This is different from therapeutic plasma exchange (plasma discarded and replaced).
Not included: Plasmaflo, Cascadeflo, Rheofilter and other membrane separators/fractionators, the Plasauto console, or HEMOSORBA CHS; these are not the separated-plasma adsorber columns covered here. This guide provides no whole-blood sorbent prescription. [1,2]
Asahi plasma-only adsorber models
Tryptophan bound to polyvinyl-alcohol (PVA) gel; immunoadsorption of selected autoantibodies and immune complexes.
Media: 350 mLColumn priming: 300 mL
Single-use [3,5]
Phenylalanine bound to PVA gel; immunoadsorption in selected rheumatologic / autoimmune pathways.
Media: 350 mLColumn priming: 300 mL
Single-use [4,5]
Styrene–divinylbenzene adsorbent for bilirubin and bile-acid removal from plasma.
Media: 350 mLColumn priming: 130 mL
International listing [6]
Porous anion-exchange resin for bilirubin and bile-acid adsorption on separated plasma.
Media: 300 mLColumn priming: 110 mL
Japan-market listing [2]
| Model | Primary adsorption target | Relevant documented applications | Key verification |
|---|---|---|---|
| IMMUSORBA TR-350 | Selected pathogenic antibodies (including anti-acetylcholine-receptor antibodies) and immune complexes | Autoimmune neurological diseases: myasthenia gravis, GBS and selected CIDP pathways | ACE inhibitor exclusion; plasma-only route; low-platelet plasma; approved IFU [3,8] |
| IMMUSORBA PH-350 | Autoantibodies and immune complexes through phenylalanine ligand | Selected autoimmune / rheumatologic diseases, including SLE and lupus nephritis in publications | ACE inhibitor exclusion; plasma-only route; applicable indication [4,9] |
| PLASORBA BR-350 (L) | Bilirubin and bile acids | Selected severe hyperbilirubinemia/cholestatic and liver support cases | Manufacturer caution: plasma only, no platelet-rich plasma; exact column/filter IFU [6,10] |
| PLASORBA BRS-350 | Bilirubin and bile acids | Japanese product listing, including fulminant hepatitis and postoperative liver failure | Japan-specific labeling, 300 mL resin vs BR-350 350 mL, market approval [2] |
The sorbent media and column priming volumes are device specifications, not plasma dose recommendations. Additional tubing, separator and protective-filter priming volumes increase total extracorporeal volume. [2–6]
Indications and therapeutic endpoints
Manufacturer clinical examples include myasthenia gravis, Guillain–Barré syndrome, CIDP and multiple sclerosis. For each diagnosis, distinguish a clinical report or product information from a guideline-endorsed first-line treatment.
Possible endpoints: clinical neurological recovery, disease-specific scale, target-antibody levels where useful, reduced rescue therapy requirement.
Manufacturer examples include systemic lupus erythematosus, malignant rheumatoid arthritis and selected neurologic diseases. Published applications include refractory lupus nephritis.
Possible endpoints: disease activity, organ function, antibody/immune complex trend, need for concomitant immunosuppression.
Selective adsorption of bilirubin and bile acids in selected severe hyperbilirubinemia or liver-support pathways. Assess underlying liver disease and urgency of transplantation/specialist care.
Possible endpoints: total/direct bilirubin change, clinical pruritus or cholestasis signs, liver and renal function, organ-support decisions.
ASFA 2026 categories apply to disease indications and procedures, not automatically to every brand or disease listed in a catalogue. Confirm the exact indication and local reimbursement/approval before prescription.
Review available alternatives including TPE, IVIg and disease-modifying treatment as appropriate. [11]
Separated-plasma adsorption pathway
The adsorber processes the separated plasma fraction, not the cellular blood stream. The complete circuit may include product-specific particle-removal filter, air traps and pressure sensors. The plasma separator and console are circuit elements, not additional sorbents.
- Single-use immunoadsorption: IMMUSORBA TR/PH use dedicated columns for a session, without assuming online regeneration.
- Bilirubin adsorption: PLASORBA BR/BRS uses a qualified plasma-separation path with plasma return, without routine full-volume donor plasma or albumin replacement.
- Do not interchange: BR-350 and BRS-350 have different documented cartridge/priming specifications and market-specific instructions.
Asahi Kasei specifically warns that its TR, PH and BR columns are designed for plasma; platelet-rich plasma may cause blockage. Check any plasma filtration alarms and platelet contamination per local protocol. [3,4,6,12]
What must be specified before each session
| Prescription field | Clinical approach | Evidence or limitation |
|---|---|---|
| Adsorber model | TR-350 / PH-350 / BR-350 (L) / Japan-market BRS-350 | Check exact label, regulatory indication and lot/expiry [1–6] |
| Separator and machine | Specify plasma separator, blood circuit, plasma pump, pressure sensors, downstream filter and disposables | Compatibility must be documented in current IFUs |
| Vascular access | Choose access allowing stable, safe extracorporeal flow | Depends on patient and machine |
| Blood flow (Qb) | Prescribe an individualized range and safety alarm thresholds | No universal Qb authorized for all four products |
| Plasma flow (Qp) | Specify per separator extraction fraction, hematocrit, adsorber capacity and pressures | Old TR-350 clinical study reported 20–30 mL/min (example only) [13] |
| Processed plasma target | Write target volume (L) and stop rule; distinguish from plasma removed | Published TR-350 studies used ~2 L in specific contexts; not a general dose [8,13] |
| Session duration | Calculated from intended Qp and target processed volume, subject to capacity and safety limits | Individualize per clinical goals and IFU |
| Number and frequency | Document disease-specific course and reassessment timetable | Cannot be inferred from one manufacturer's cartridge size |
| Replacement/fluid balance | Processed plasma normally returned; chart priming, net balance and replacement for losses if needed | Not the same as TPE replacement of exchanged plasma |
| Anticoagulation | Approved systemic or regional protocol, laboratory targets, calcium strategy if citrate | No universal heparin/citrate dose for these columns |
Processed plasma estimator
Estimate the amount of separated plasma passing through the column. This calculator is a planning aid only; it does not determine the clinical dose, approved flow or treatment duration.
Illustrative calculation only.
Column information will update with your selection.
Estimated plasma volume, when needed for comparison, uses the simplified formula EPV (L) = 0.065 × weight (kg) × (1 − hematocrit fraction). No model-specific validated treatment target is inferred from this ratio.
Pre-treatment and circuit setup
- Confirm the patient's indication, prior therapies, clinical stability, consent and appropriate specialist team.
- Verify adsorber specific model, country-specific IFU, expiry and sterile package integrity; plan the compatible separator, tubing set, machine mode and particle-removal filter where indicated.
- Review medications, especially ACE inhibitors before TR-350/PH-350, and any drug potentially removed by adsorption.
- Document extracorporeal circuit/adsorber priming fluid and volume; perform manufacturer-prescribed priming and air removal; do not improvise rinse recipes.
- Ensure the separated plasma is free of excessive platelets before it enters the adsorber; do not pass whole blood through these columns.
- Prescribe anticoagulant agent and monitoring according to the patient's coagulation/bleeding risk and approved system protocol. For citrate, review calcium, acid–base status and organ function.
- Record baseline pressures, alarms, blood/plasma flow targets, fluid balance and emergency bypass/stop actions.
Clinical and laboratory surveillance
| Timing | Required considerations |
|---|---|
| Before treatment | Vitals, volume status, access, medications, CBC/platelets, electrolytes; calcium and coagulation according to anticoagulation; indication-specific neurologic/autoimmune markers or liver profile and bilirubin |
| During session | Blood pressure, pulse, temperature and symptoms; blood/plasma flow; separator and sorbent pressures; clotting/air alarms, anticoagulation tolerance and fluid balance; severe hypotension/reaction surveillance |
| At completion | Clinical tolerance and circuit return, bleeding/hemolysis signs; review selected lab values and treatment goal; documentation of actual treated volume and adverse events |
| Between sessions | Functional neurologic response / disease-specific scores, antibody titers where informative, organ function; for bilirubin adsorbers total/direct bilirubin and liver clinical course; reassess benefit vs harm |
Rising adsorber pressure, poor plasma separation, suspected platelet breakthrough or clotting requires evaluation under the device's stop/bypass protocol. Drug exposure may be altered by selective plasma adsorption, so consider medication timing and relevant therapeutic drug monitoring.
Safety-critical review
- Hypotension, including ACE inhibitor–associated reactions for TR/PH
- Thrombotic or blocked circuit / abnormal pressures
- Allergy or biocompatibility reaction
- Bleeding, platelet or coagulation complications
- Catheter infection or access problems
- Citrate-related hypocalcemia or acid–base disturbance
- Severe hemodynamic instability or anaphylactoid symptoms occur
- Air / flow / pressure alarms remain unresolved
- Clotting, plasma leak or hemolysis is suspected
- Bleeding or serious laboratory deterioration develops
- Benefit is absent or risk exceeds the predefined target
Selection at a glance
| Feature | TR-350 | PH-350 | BR-350 (L) | BRS-350 |
|---|---|---|---|---|
| Route | Separated plasma | Separated plasma | Separated plasma | Separated plasma |
| Technology | Tryptophan–PVA | Phenylalanine–PVA | Styrene–divinylbenzene resin | Porous anion-exchange resin |
| Target | Selected autoantibodies / immune complexes | Selected autoantibodies / immune complexes | Bilirubin / bile acids | Bilirubin / bile acids |
| Adsorbent media | 350 mL | 350 mL | 350 mL | 300 mL |
| Column priming | 300 mL | 300 mL | 130 mL | 110 mL |
| Regional source | Global + Japan | Global + Japan | International Asahi catalogue | Japan professional catalogue |
| Key constraint | ACE inhibitors | ACE inhibitors | Platelet-rich plasma | Japan-specific IFU |
Product dimensions and priming volumes are approximate catalogue figures for the specified variants, not requirements to use a particular session time or dose. [2–6]
Asahi plasma adsorption prescription checklist
Documentation framework only. Final medical orders require patient-specific qualified clinical review and current approved manufacturer instructions.
Prescription Worksheet — Complete & Print
Fill in this patient-specific prescription worksheet on screen, copy the completed fields, or print the worksheet alone. Blank fields are intentional; the worksheet does not determine appropriate device settings or constitute an authorized medical order.
Mandatory clinical verification: the signed institutional order, current model-specific manufacturer IFU, applicable regulatory authorization and patient-specific medical review take precedence. The worksheet remains entirely local to this page; its entries are not saved or transmitted.
Manufacturer references and evidence sources
- Asahi Kasei Medical — international apheresis product line. Current product listing separating plasma adsorbers from membrane separators and equipment.
- Asahi Kasei Medical Japan — PLASORBA BRS product listing. BRS-350 Japan-market model, ion-exchange resin, 300 mL media, 110 mL priming; local regulatory reference.
- IMMUSORBA TR-350 — manufacturer specifications and cautions. 350 mL tryptophan–PVA media; 300 mL priming; plasma-only and ACE-inhibitor warning.
- IMMUSORBA TR/PH — Japan professional product listing. Both 350 mL PVA columns and distinct tryptophan/phenylalanine ligands; Japan approvals.
- IMMUSORBA PH — international product information. Selected autoimmune uses; confirm exact local IFU.
- PLASORBA BR-350 — manufacturer specifications and cautions. 350 mL adsorbent media, 130 mL priming and plasma-only restriction.
- Japan PMDA Medical Safety Information No. 68 (2024). Japanese regulator warning about ACE inhibitors and specified immunoadsorption columns.
- IMMUSORBA TR — manufacturer clinical performance examples. Includes selected treated-volume and disease-response observations.
- Asahi Kasei — bibliographic references by product. Publications for IMMUSORBA PH and TR, and PLASORBA BR.
- PLASORBA BR — product indication overview. Bilirubin and bile acid removal, liver-support examples.
- ASFA 2026 — 10th Special Issue therapeutic apheresis guidelines. Indication-level recommendation categories and evidence grades, not proprietary device settings.
- Asahi Kasei — mechanism of plasma adsorption and immunoadsorption. Conceptual plasma separation and adsorption pathway.
- Plasmaperfusion on tryptophan columns in myasthenia gravis. Historical study-specific example of plasma flow and treated volume. Not an operating manual.