1. Definition and therapeutic objectives
Whole-blood hemoadsorption (HA), also called direct hemoperfusion (HP) when blood traverses a sorbent cartridge, is an extracorporeal technique in which unseparated blood flows through adsorptive media and is returned to the patient. Targets depend on the cartridge's ligand, pore size, surface chemistry and binding capacity. [2]
- Potential targets: inflammatory mediators, endotoxin, selected endogenous/exogenous toxins, circulating pathogens, lipoproteins or fibrinogen—not interchangeable between devices.
- Routes: a dedicated hemoperfusion circuit or a validated side/inline connection to CRRT, cardiopulmonary bypass (CPB), or ECMO.
- No mandatory plasma separator: the blood cellular fraction passes through the adsorber; there is normally no planned exchange of plasma with replacement fluid.
- Endpoint: a defined clinical response and/or target-specific biochemical response; the volume of blood pumped through a column is not equivalent to toxin clearance.
2. Circuit configuration and modalities
Standalone hemoperfusion
Blood is drawn from vascular access, pumped through a compatible adsorber and returned to the patient. There is no plasma-separation step and no prescribed plasma replacement.
Use only a validated machine, tubing set, pump and anticoagulation configuration.
Combined therapy
The cartridge may be placed in a manufacturer-validated blood path associated with CRRT, intermittent HD, CPB or ECMO. Position, bypass, direction of flow, pressure and alarm logic are system-specific.
An approved device combination does not imply all placements or line arrangements are safe.
| Configuration | Prescription considerations |
|---|---|
| Standalone direct HP | Specify blood pump, vascular access, extracorporeal volume, blood flow, hemoadsorber pressure limits, anticoagulation, fluid balance and blood return plan. |
| With CRRT or HD | Identify exact upstream/downstream position per manufacturer instructions; coordinate blood flow, anticoagulation, filtration and prescribed kidney replacement treatment. |
| With ECMO / CPB | Only a manufacturer-validated connection arrangement; confirm bypass and line-pressure conditions, no unauthorized changes to the main flow, air safety and team responsibilities. |
| Direct hemoadsorption vs plasma adsorption | Do not substitute a plasma-only column into a whole-blood circuit, or vice versa, without explicit model-specific authorisation. |
3. Core prescription parameters
Every prescription must document the exact cartridge and extracorporeal system. There is no safe universal Qb, duration, exchange interval, or anticoagulant dose for all hemoadsorption devices. [2]
| Parameter | Required physician / team decision | Verification |
|---|---|---|
| Clinical indication | Diagnosis, severity, reason for using HA rather than standard care alone, proposed response measure | Current evidence, local authorization and multidisciplinary approval |
| Cartridge model | Brand, exact model/catalogue number, target class, lot, expiry, single-use status | Current market-specific device IFU |
| Circuit / machine | Standalone HP / CRRT / HD / ECMO / CPB; insertion position and compatible disposables | Machine + cartridge IFUs, engineering checks |
| Prescribed blood flow (Qb) | Patient- and device-specific target, in mL/min | Access capacity, cartridge pressure-drop curve and alarms |
| Active therapy duration | Planned hours, intended number of cartridges, stopping/reassessment criteria | Model-specific upper treatment time and saturation constraints |
| Cartridge change | Trigger by IFU limit, pressure/clotting or predefined clinician reassessment | Not assumed from another manufacturer's lifespan |
| Anticoagulation | Agent, patient-specific regimen, monitoring and HIT/bleeding precautions | Compatibility with the full circuit |
| Fluid management | Priming/rinsing, rinse-back, net balance, extracorporeal blood volume | Manufacturer procedures; hemodynamic tolerance |
| Drug therapy | Antimicrobial, anticonvulsant, anticoagulant and other relevant exposures | Pharmacy review / therapeutic drug monitoring (TDM) as available |
| Monitoring | Clinical endpoints, cartridge pressure and delivered flow, platelets/coagulation, labs | Baseline, on-treatment and post-treatment documentation |
4. Processed blood volume — Interactive calculator
This tool calculates the volume of whole blood passing through the cartridge, not plasma volume and not an adsorptive dose, target concentration reduction or drug clearance. No numerical inputs are preselected.
Blood-processing estimator (illustrative only)
Enter measured blood flow and duration to estimate volume.
5. Adsorber selection — Distinct targets
The following are examples of whole-blood-capable sorbent families, not a substitution or interchangeability table. Verify exact model, intended-use route and local regulatory status. Device claims below reflect manufacturer documentation and may exceed the strength of patient-outcome evidence.
| Cartridge / manufacturer | Documented whole-blood adsorption target or purpose | Clinical / safety distinction |
|---|---|---|
| CytoSorb — CytoSorbents | Broad adsorption of selected hydrophobic middle molecules / inflammatory mediators | Can be integrated into validated extracorporeal circuits; drug removal and cartridge change timing need separate review. [3] |
| HA series (HA130 / HA230 / HA330 / HA380) — Jafron | Resin hemoperfusion; model-specific toxins and mediator classes | Avoid transferring a HA-family prescription to BS330, which is a plasma-route bilirubin adsorber. [5] |
| MG series — Biosky / Biosun | Whole-blood hemoperfusion resin for medium/large molecular substances | MG is distinct from DX plasma bilirubin adsorption. [8] |
| TORAYMYXIN PMX-20R / PMX-05R — Toray Medical | Polymyxin-B-immobilized endotoxin adsorption | Selective endotoxin device, with model-specific circuits; SSC 2026 suggests against routine PMX hemoperfusion for sepsis. [1] [4] |
| Alteco LPS Adsorber — Alteco Medical | Endotoxin adsorption using a synthetic peptide ligand | Manufacturer-specific priming, blood flow and treatment time; outcomes remain clinically uncertain. [6] |
| Seraph 100 — ExThera Medical | Heparin-mimetic affinity adsorption of certain pathogens | Regulatory indications are jurisdiction-specific; U.S. historical emergency-use authorization does not equal unrestricted approval. [7] |
| DALI 500 / DALI 750 — Fresenius Medical Care | Direct adsorption of LDL-C and Lp(a) from whole blood | Lipoprotein apheresis indication, not a sepsis/cytokine device. [9] |
| Rheocarna — Kaneka Medix | Whole-blood adsorption targeting LDL cholesterol and fibrinogen | Japan-specific refractory-ulcer indication; verify local distribution/authorization. [10] |
6. Preparation, vascular access and anticoagulation
Pre-connection checklist
- Confirm exact model, sterile packaging, expiry, single-use indication and IFU revision.
- Check access patency, suitable blood-pump system, line compatibility, adsorption-column orientation and air-management configuration.
- Document extracorporeal blood volume, priming/rinsing instructions, saline/rinse-back and potential hemodynamic effects.
- Validate planned pressures, alarms, delivered flow monitoring and emergency bypass / blood return procedures.
Anticoagulation plan
- Select systemic heparin, regional citrate (where permitted), or another validated protocol according to the entire circuit and patient risks.
- Check bleeding risk, thrombocytopenia, prior HIT, liver dysfunction and citrate compatibility.
- Monitor according to the chosen approach: ACT/aPTT or other local measures; ionized calcium / acid–base status when citrate is used.
- Do not infer safe anticoagulation dosage from general hemoadsorption product brochures.
7. Clinical, technical and pharmacological monitoring
| Time | Required observations / checks |
|---|---|
| Before each session | Vital signs, hemodynamic support, indication-specific biomarkers, access; CBC/platelets, coagulation and electrolytes; medication review and baseline drug levels where relevant. |
| During treatment — clinical | Blood pressure, heart rate/rhythm, temperature, symptoms, vasopressor support, bleeding or hypersensitivity, net fluid balance. |
| During treatment — technical | Delivered Qb, circuit pressure trends and pressure drop across the cartridge, clotting/air alarms, elapsed active duration, circuit position and delivered anticoagulation. |
| Drug safety | Review antibiotic and antifungal dosing, antiepileptics, anticoagulants and other high-risk drugs; TDM as available. Consider removal caused by concurrent CRRT separately. |
| End of session | Actual treatment time and blood volume processed, cartridge changes, reason for discontinuation, blood return, adverse events and procedure-related losses. |
| Course / follow-up | Prespecified clinical outcome (not only biomarker reduction), organ support, laboratory trends, thrombocytopenia, rebound, medication exposure, decision to continue or stop. |
Document any interpretation challenges caused by adsorption of biomarkers as well as adsorption of the intended target. [2]
8. Clinical applications and evidence boundaries
| Setting / target | How to assess use | Evidence caveat |
|---|---|---|
| Sepsis / septic shock — cytokines or endotoxin | Identify clinical rationale, established source control and standard sepsis bundle, clear stop endpoints and local research/approval framework. | SSC 2026 suggests against blood purification including hemoperfusion and against polymyxin-B hemoperfusion; not a routine adjunct. [1] |
| Drug overdose or poisoning | Substance, toxicokinetics, protein binding, alternative extracorporeal removal and toxicology consultation. | Only use validated sorbent/route when toxin adsorption is plausible and standard antidotes/support remain prioritized. [2] |
| ESRD-related symptoms / toxin adsorption | Disease-specific assessment and compatible HD+HA prescription when applicable. | Evidence and local practice vary; no general survival or symptom-benefit claim. [12] |
| Direct lipoprotein adsorption | Document severe lipid disease or vascular/ulcer indication and multidisciplinary assessment. | Cartridge-specific indications, separate from inflammatory hemoadsorption. [9] [10] |
| Pathogen affinity adsorption | Microbiology, bacteremia or infection phenotype, approval status and antimicrobial plan. | Potential pathogen removal does not imply established outcome benefit or unrestricted regulatory approval. [7] |
| Perioperative extracorporeal support | Determine if insertion is supported by combined circuit IFUs; monitor coagulation and drug exposure. | Pilot/observational evidence does not establish routine benefit across operations. [2] |
9. Adverse events and stopping rules
| Risk | Practical response |
|---|---|
| Hypotension / hemodynamic instability | Assess blood pressure, fluid shifts, access, circuit resistance and possible acute reaction; initiate institutional emergency response. |
| Clotting, high pressure, air or circuit failure | Stop/escalate according to machine and cartridge IFUs; never override safety alarms to maintain treatment. |
| Thrombocytopenia, leukocyte or protein loss | Monitor CBC/platelets, bleeding and other relevant tests; reassess risk–benefit, particularly with repeated use. |
| Hypocalcemia / citrate accumulation | Monitor ionized calcium and acid–base balance where citrate is approved; clinician-led correction per protocol. |
| Hypersensitivity / biocompatibility reaction | Immediate recognition and emergency management per institution; check device contraindications. |
| Drug underexposure | Urgent pharmacy review, dose/timing adjustment and TDM when feasible; especially anti-infectives and other critical medicines. |
| No clinical benefit / objective achieved | Discontinue if predefined endpoint is met or treatment is futile; reassess on patient-centered outcomes, not only biomarker levels. |
10. Standardized hemoadsorption prescription template
Structured documentation items for the prescribing team. The blank, fillable worksheet follows in Section 11 and can be printed by itself.
| Required field | What must be recorded |
|---|---|
| Indication / evidence | Disease, treatment objective, care pathway and clinician authorization |
| Exact adsorber | Brand, catalog/model number, ligand/target and regulatory/IFU confirmation |
| Combined circuit | HP alone vs CRRT / HD / CPB / ECMO, approved connection point and machine |
| Access / blood flow | Vascular access, prescribed Qb and validated pressure/alarm limits |
| Planned processing | Active duration, planned cartridge count and device-specific change conditions |
| Anticoagulation / fluids | Agent and monitoring, priming, circuit volume and return plan |
| Medication/TDM plan | Drugs at risk of adsorptive clearance, level checks and timing |
| Response / stop rules | Clinical and technical endpoints, escalation and stopping criteria |
| Safety / sign-off | Monitoring plan, treatment authorization, responsible clinician, signature and date |
This framework is educational and cannot be used as a standing or executable medical order.
Prescription Worksheet — Complete & Print
Complete this patient-specific whole-blood hemoadsorption worksheet on screen, copy its contents, or print it separately. Blank fields are deliberate: no cartridge setting, flow, dose or anticoagulant is automatically prescribed.
Mandatory clinical verification: This worksheet does not select a cartridge, prescribe Qb or define anticoagulation. The signed institutional order, current manufacturer IFUs and authorisation, clinical review and local protocol take precedence. Entries stay in this local browser page and are not transmitted or stored by this document.
12. Key references and practice summary
- Surviving Sepsis Campaign, 2026. International Guidelines for Management of Sepsis and Septic Shock — suggests against blood purification (including hemoperfusion) and polymyxin-B hemoperfusion in adults with sepsis/septic shock. Read source.
- ADQI 30 Consensus, 2024. Hemoadsorption: consensus report of the 30th Acute Disease Quality Initiative workgroup. Nephrology Dialysis Transplantation 39:1945–1964. Technology, indications, uncertainty, prescribing, monitoring and risks. Read source.
- CytoSorbents — CytoSorb technical guide. Manufacturer-specific setup, blood flow, anticoagulation context and maximum use time. These are not general hemoadsorption settings. Read source.
- Toray Medical — TORAYMYXIN. Manufacturer product description and differentiated PMX models for endotoxin adsorption. Read source.
- Jafron Biomedical — HA series. Manufacturer product portfolio describing direct whole-blood hemoperfusion HA cartridges and separate plasma BS columns. Read source.
- Alteco Medical — LPS Adsorber. Direct blood endotoxin adsorption product and technical setup information. Read source.
- ExThera Medical — Seraph 100. Manufacturer overview of pathogen-affinity blood adsorption; regulatory status is jurisdiction-specific. Read source.
- Biosun — MG disposable hemoperfusion cartridge. Whole-blood MG series mechanism and advertised areas of application; distinguish from plasma DX. Read source.
- Fresenius Medical Care — DALI instructions for use. DALI 500/750 direct whole-blood lipoprotein adsorption; safety and approval are model-specific. Read source.
- Kaneka Medix — Rheocarna. Manufacturer product page for LDL-cholesterol and fibrinogen whole-blood adsorption in a Japan-specific refractory-ulcer indication. Read source.
- Systematic drug-clearance review, 2025. Current Knowledge on the Impact of Extracorporeal Blood Purification Devices on Drug Clearance — risk of clinically relevant drug removal and importance of TDM. Read source.
- International expert consensus on hemoadsorption with hemodialysis, 2026. Hemoadsorption combined with hemodialysis: a consensus statement — selected maintenance-HD configurations, evidence limitations and drug monitoring. Read source.
Essential prescribing principles
Define the clinical purpose and evidence boundary; choose the exact whole-blood adsorber and compatible circuit; document Qb, actual active duration, device-specific change limits, anticoagulation, drug/TDM plan, safety monitoring and stopping criteria. No plasma volume multiplier or generic cartridge schedule should be transferred from immunoadsorption, DFPP or plasma adsorption.