1. Scope & evidence boundaries
Covered: whole-blood extracorporeal circuit safety; therapeutic plasma exchange by centrifugation or membrane filtration; plasma adsorption; immunoadsorption; whole-blood hemoadsorption; double-filtration plasmapheresis (DFPP); cascade filtration/rheopheresis; and extracorporeal photopheresis (ECP). Not covered: hemodialysis or renal replacement anticoagulation protocols.
Terminology: “whole blood” below means safety of an extracorporeal blood circuit, not a stand-alone treatment indication. DFPP and cascade filtration commonly describe the same two-filter principle; they have separate navigation entries for practical usability, not because they require fundamentally different risk protocols. [R1] [R2] [R12]
Hemodynamic status, bleeding risk, anticoagulant reactions, allergy and access injury
Validated blood path, pressure limits, air detection, hemolysis and clot formation
Plasma protein loss, sorbent interactions, drug removal, cell depletion or methoxsalen exposure
2. Technique-specific safety matrix
Choose one technique using the selector above. The matrix, navigation and detailed safety chapter will change together. Shared triage and worksheet sections remain visible for every choice.
| Technique | Principal complications | Priority safeguards | Supporting evidence |
|---|---|---|---|
| Whole-blood extracorporeal circuit | Air/clot/hemolysis; access injury; unexpected blood loss | Trace and prime; pressure/air alarm checks; trained return-of-blood procedure | [R2] [R4] |
| TPE — centrifugation | Citrate hypocalcemia; plasma allergy; hypotension; coagulation-factor depletion | Ionized Ca and symptoms; blood-product checks; volume and fibrinogen plan | [R2] [R8] |
| TPE — membrane filtration | Transmembrane pressure rise; filter clotting/hemolysis; hypotension; coagulopathy | Monitor TMP/hemolysis; approved circuit anticoagulation; fluid/coagulation plan | [R6] [R7] |
| Plasma adsorption / DPMAS | Bleeding or coagulopathy; hypotension; allergic reaction; cartridge clotting | Platelets/fibrinogen/INR; pressures; access and fluid monitoring | [R12] [R15] |
| Immunoadsorption | Adsorber/clotting or allergy; antibody/protein depletion; catheter events | Check ligand and regeneration IFU; Ig targets; access and infection surveillance | [R12] [R13] [R14] |
| Whole-blood hemoadsorption | Thrombocytopenia; drug removal; pressure/clotting; hypersensitivity | CBC and medication review; pressures; validated cartridge/anticoagulation | [R16] [R17] |
| DFPP | Secondary-filter pressure rise; hemolysis; fibrinogen/Ig loss; hypotension | TMP1/TMP2 trend; coagulation/fibrinogen; albumin and balance assessment | [R9] [R10] |
| Cascade filtration / rheopheresis | DFPP-type circuit risks; target-dependent protein depletion | Verify secondary-filter pore selectivity, pressure and replacement plan | [R11] [R21] |
| Extracorporeal photopheresis | Hypotension; cytopenia; fever; drug/light sensitivity | Extracorporeal volume; CBC; psoralen screening; platform IFU | [R18] [R19] [R20] |
3. Pre-procedure clinical & technical safety screen
Before treatment
- Confirm indication, rationale, expected benefit and consent; identify special vulnerability (hemodynamic instability, allergy, bleeding, pregnancy and low body weight where relevant).
- Review CBC/platelets, PT/INR, aPTT, fibrinogen, electrolytes (including ionized Ca and Mg when citrate may be used), renal/liver function where clinically relevant.
- Check allergy to albumin/plasma, anticoagulants, membrane/sorbent and psoralen where relevant; assess previous HIT and ACE-inhibitor exposure.
- Choose access and confirm catheter position/patency, asepsis, compatible machine/kit/lot and approved anticoagulation strategy.
- Document fluid/replacement, medication timing, blood conservation and contingency plans.
During treatment
- Baseline and interval vitals, symptoms, fluid balance, access and circuit pressure/air alarms; observe the patient—not alarms alone.
- Check for citrate symptoms, bleeding, anaphylaxis, clotting, hemolysis, and target-specific losses.
- Escalate abnormal trends according to the local protocol; obtain urgent laboratory tests when clinically indicated.
After treatment
- Recheck clinical state, access site, fluid balance and indicated CBC/coagulation/electrolytes.
- Document actual treatment delivered, product exposure, adverse events, medication timing and a plan for subsequent sessions.
- Report serious or device-related events to the institutional safety team and manufacturer as appropriate.
A system-wide safety check cannot replace the technique-specific IFU. [R2] [R3] [R4]
4. Whole-blood extracorporeal circuit — common hazards
Applies to any procedure that withdraws and returns whole blood through tubing, pumps and vascular access. This is a cross-cutting circuit safety category, not a separate blood purification treatment. [R2] [R4]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Vascular access bleeding, hematoma, thrombosis or infection | Oozing, swelling, difficult aspiration, pain, erythema, fever | Aseptic access, secure dressings, verify patency and anticoagulation plan; reassess access each session | Pause or interrupt as appropriate; assess bleeding/thrombosis/infection; escalate urgently if major event. |
| Air entry / embolism | Air alarm, air in tubing, abrupt dyspnea, chest symptoms or collapse | Validated air detectors, intact connections, correct priming and alarm checks | Stop per emergency circuit protocol; initiate immediate clinical emergency response; return of circuit blood depends on event and IFU. |
| Clotting / rising circuit pressure | Visible clots; pressure increases; flow failure | Correct anticoagulation, air removal, pressure monitoring and allowed flow configuration | Do not bypass alarms; stop or replace circuit per IFU. Do not return potentially compromised circuit blood without clinical/device authorization. |
| Hemolysis / mechanical blood trauma | Dark plasma, red/brown effluent, pain, unexplained K rise or instability | Check tubing kinks, inappropriate pressures and machine/pump setup | Urgently halt suspected causative circuit and assess hemolysis, electrolytes and clinical stability. |
| Extracorporeal blood loss, hypovolemia, hypothermia | Hypotension, tachycardia, cold extremities, low return volume | Document extracorporeal volume and returned blood; warming/fluid plan where appropriate | Assess and support perfusion, temperature and fluid balance; decide continuation after reassessment. |
5. TPE by centrifugation — complications & prevention
Centrifugal therapeutic plasma exchange commonly uses citrate at the blood inlet; reaction patterns are strongly influenced by replacement fluid and total citrate exposure. [R2] [R5] [R8]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Citrate-associated hypocalcemia ± hypomagnesemia | Perioral tingling, cramps, nausea; severe tetany, QT prolongation or arrhythmia | Screen risk factors; monitor symptoms and ionized Ca (plus Mg where indicated); validate citrate/calcium regimen and source of citrate in donor plasma | Follow local citrate-toxicity pathway: reassess procedure/citrate exposure and treat symptomatic electrolyte disturbance urgently. |
| Hypotension or vasovagal reaction | Dizziness, sweating, low BP, nausea or reduced consciousness | Hemodynamic risk screen; appropriate replacement balance and extracorporeal-volume planning | Pause procedure and assess volume status, allergic causes, arrhythmia and other shock causes; escalate. |
| Allergy / anaphylaxis / transfusion reaction | Rash, wheeze, flushing, fever or collapse after plasma exposure | Verify identity and plasma product; screen allergy history; have reaction protocol and resuscitation equipment | Stop exposure and activate institutional transfusion/anaphylaxis response; preserve product data for investigation. |
| Coagulopathy and loss of protective plasma constituents | New bleeding, fibrinogen fall, prolonged coagulation testing | Baseline and repeat fibrinogen/platelets/coagulation tests in high-risk or repeated sessions; individualize replacement strategy | Reassess further exchanges and blood-product needs with treating physician. |
| Medication removal or interactions | Lower-than-expected drug effect or level after exchange | Review protein binding, distribution volume and dose timing of critical medications | Use therapeutic drug monitoring or specialist pharmacist plan when indicated. |
Before treatment
- Confirm citrate and replacement-fluid plan; allergy, coagulation and calcium baseline.
During treatment
- Assess neurologic/cardiac symptoms and vitals; check iCa at locally required intervals; observe for transfusion reactions.
After treatment
- Examine puncture sites; reassess fibrinogen/electrolytes when indicated; review drug timing and next treatment.
6. TPE by membrane filtration — complications & prevention
Primary membrane plasma separation has particular concerns about transmembrane pressure, progressive filter clotting and mechanical hemolysis. Some filters may increase bradykinin-related risk in patients exposed to ACE inhibitors. [R6] [R7] [R12]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Excess transmembrane pressure / filter clotting | TMP increase, falling plasma flow, alarm or visible clot | Use approved plasma fraction and blood/plasma flow ranges; correct anticoagulation and pressure trend monitoring | Do not increase pressure past IFU limits. Investigate and replace circuit/filter when indicated. |
| Hemolysis | Hemolyzed effluent, dark plasma, unexpected pain, instability or electrolyte change | Avoid excessive TMP, pressure gradients and tubing obstruction; monitor color and circuit data | Stop suspected harmful circuit; promptly evaluate hemolysis and cardiac/electrolyte risk. |
| Hypotension / fluid imbalance | BP fall, poor perfusion, tachycardia | Plan replacement/fluid balance, blood-flow tolerance and access compatibility | Assess cause and restore hemodynamic stability per institutional protocol. |
| Coagulation-factor/fibrinogen depletion | Bleeding or low fibrinogen after plasma removal | Follow CBC, fibrinogen and coagulation tests based on patient risk and repeat exchanges | Reassess replacement and continuation before further exchange. |
| Bradykinin-associated reaction (selected membranes / ACE inhibitors) | Abrupt flushing, hypotension, dyspnea | Review ACE inhibitor exposure and the exact membrane IFU; follow any required drug interruption window | Treat as a severe reaction until proven otherwise; stop the implicated system and escalate. |
| Replacement-plasma reactions or citrate exposure | Allergy, respiratory symptoms or tingling during plasma replacement | Product verification; citrate and calcium awareness even if the circuit uses heparin | Follow transfusion-reaction and electrolyte evaluation pathways. |
7. Plasma adsorption / bilirubin adsorption / DPMAS
In plasma adsorption, separated plasma passes through one or more sorbent columns and returns to the patient. Risk depends on the ligand/resin, cartridge capacity, anticoagulant and the patient’s underlying coagulopathy; DPMAS liver-failure populations need special vigilance. [R12] [R15]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Bleeding / worse coagulation | Puncture bleeding, mucosal bleed, falling fibrinogen or platelets, rising INR | Document baseline CBC, coagulation/fibrinogen and repeat with clinical severity; assess adsorption-related and liver-related losses | Reassess anticoagulation and procedural continuation; manage bleeding under local protocol. |
| Hypotension / extracorporeal-volume intolerance | Early BP drop, tachycardia, dizziness | Pre-treatment hemodynamic assessment, appropriate priming and careful fluid plan | Pause and assess perfusion, volume, allergic and bleeding causes. |
| Allergy / biocompatibility reaction | Flushing, urticaria, bronchospasm | Confirm resin/ligand history, priming/wash and approved materials | Terminate suspected triggering exposure and treat reaction urgently if severe. |
| Column clotting / pressure rise | Inlet/outlet pressure changes, loss of plasma flow | Validated anticoagulant and cartridge orientation; trend pressures, respect cartridge lifespan | Stop or replace malfunctioning column per IFU; do not bypass alarms. |
| Unintended protein or medication adsorption | Unexpected laboratory/protein decline or loss of drug effect | Review target and potential off-target binding; monitor clinically critical proteins/drug concentrations where feasible | Consider dose timing or adjustment with pharmacy/clinical team; do not assume selective adsorption is perfectly specific. |
Before treatment
- Name the exact sorbents and target, separator, coagulation status, anticoagulation and emergency limits.
During treatment
- Monitor access, separated-plasma flow, column pressures, hemodynamics and bleeding.
After treatment
- Document delivered plasma volume, pressure events, blood loss, CBC/coagulation changes and medication review.
8. Immunoadsorption (including ABO-specific)
Immunoadsorption removes selected antibodies/immune complexes after plasma separation. Single-use and regenerable columns have different safety steps. ABO-specific columns also require target-antibody monitoring, and the overall safety assessment depends on the transplant or disease context. [R1] [R13] [R14]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Bradykinin-related severe hypotension in at-risk circuits | Sudden flushing, dyspnea, hypotension, bradycardia | Review ACE inhibitors and the exact adsorption/filter IFU; apply manufacturer-defined restrictions | Stop the relevant exposure and manage anaphylactoid/shock presentation immediately. |
| Column-related hypersensitivity | Urticaria, airway symptoms, hypotension | Confirm ligand/material history, proper priming and approved column combinations | Manage reaction and report potential device issue; do not restart without review. |
| Excess antibody/protein depletion or immunologic vulnerability | Marked Ig reduction, infection or poor vaccine/Ig therapy timing | Track target antibodies/IgG (and other Ig classes if relevant); coordinate IVIG/biologics; assess infection risk | Adjust session plan and adjunct therapy with immunology/transplant service. |
| Regeneration/rinse or column-switching failure | Unexpected alarm, conductivity/pH change, wrong valve state | Use only validated regeneration sequence and leak/rinse safeguards; two-person verification when required | Stop device according to IFU; do not allow regeneration solution or unverified path to reach patient. |
| Access-related thrombosis and circuit clotting | Poor flow, catheter swelling/pain, access alarms, visible clot | Access assessment and anticoagulation per platform; review symptoms after multiple sessions | Evaluate suspected thrombosis and access complications promptly. |
Before treatment
- Identify ligand/ABO specificity, planned antibody monitoring, ACE-inhibitor restrictions, regeneration and rinse checks.
During treatment
- Watch for abrupt hypotension, column alarms, clotting and adverse reactions.
After treatment
- Reassess target titers and clinically relevant Ig/proteins, catheter sites and timing of IVIG/immunosuppression.
9. Whole-blood hemoadsorption / hemoperfusion
A whole-blood sorbent cartridge is perfused directly with blood. Unlike plasma adsorption, no plasma separator is necessarily present. This chapter applies to hemoadsorption circuits without introducing renal replacement treatment protocols. Cartridge chemistry, indication and integration materially affect safety. [R16] [R17]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Thrombocytopenia, anemia, leukocyte/protein loss | Platelet/Hb/WBC decline, bleeding or repeated-session blood loss | Check CBC before and after sessions as clinically indicated; document blood losses and cartridge biocompatibility | Investigate cause including underlying disease, anticoagulants and sorbent effects; reassess subsequent treatment. |
| Unintended adsorption of medicines | Drug level falls or loss of effect, e.g. time-critical antibiotic exposure | Pharmacist review of antibiotics, antiepileptics, anticoagulants and narrow-therapeutic-index drugs; use drug levels if available | Optimize dosing/timing with treating team; avoid assumptions about adsorption for a specific drug or cartridge. |
| Cartridge clotting / excessive resistance | Rising pressure gradient, flow alarms, visible clot | Confirm IFU-allowed cartridge location, priming, blood flow and anticoagulation; follow pressure limits | Stop/reconfigure only under approved procedure; do not ignore pressure alarms. |
| Hypotension or systemic reaction | BP drop, flushing, dyspnea, escalating support needs | Assess hemodynamics, volume status and compatibility; maintain emergency response readiness | Pause treatment, assess shock/anaphylaxis and activate urgent escalation if needed. |
| Blood trauma, air embolism and vascular injury | Discoloration, air alarms, catheter problems | Use whole-blood circuit checks and trained operators; trace lines and air detectors | Emergency circuit response according to IFU; evaluate clinical consequences. |
Before treatment
- Exact cartridge/indication, bleeding and cytopenia baseline; pharmacist review of essential medications.
During treatment
- Trend pressure gradient, flow, clinical stability and device time limits; assess clotting and new reactions.
After treatment
- CBC and clinically relevant drug levels; check access and medication exposure; report device-related reactions.
10. Double-filtration plasmapheresis (DFPP)
DFPP passes separated plasma through a secondary fractionation membrane to remove larger molecules; the retained fraction is discarded and the filtrate returned. The filter is not completely selective: fibrinogen, immunoglobulins and some albumin may be lost. [R9] [R10] [R11]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Secondary-filter TMP rise / occlusion | TMP2 rise, reduced permeate flow, pressure alarms | Verify secondary-filter choice, blood/plasma rates, anticoagulation and accepted pressure limits; trend both filtration stages | Do not force pressure/flow beyond IFU; investigate clotting and exchange filters as indicated. |
| Hemolysis | Dark plasma, hemolysis alarm, sudden electrolyte change | Monitor excessive TMP or shear, tubing integrity and filter compatibility | Stop suspected traumatic circuit; evaluate blood trauma and clinical impact. |
| Hypofibrinogenemia / coagulation-factor depletion | Unexpected fibrinogen drop, bleeding or prolonged coagulation tests | Use indication-specific pretreatment and repeat fibrinogen/coagulation checks; plan replacement/blood products individually | Reconsider further filtration and treat clinically meaningful factor loss per local protocol. |
| Albumin/Ig losses and fluid imbalance | Low albumin, edema, hypo-oncotic symptoms or increased infection susceptibility after repeated sessions | Quantify discarded fraction and replacement need; check proteins and antibody levels when relevant | Adjust treatment schedule, fluids and planned adjunct therapies. |
| Hypotension / access failure | BP fall, nausea, unstable flow | Risk-adapt extracorporeal volume and access; monitor continuously | Assess hypoperfusion and vascular access; interrupt as clinically appropriate. |
11. Cascade filtration / rheopheresis — specific safeguards
Cascade filtration is often a synonym for DFPP; rheopheresis denotes a filter/target-specific form used to remove selected high-molecular-weight substances. It is listed separately here so users can select “cascade filtration”, but the fundamental pressure, hemolysis, access and anticoagulation precautions overlap with DFPP. [R11] [R12] [R21]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Unintended fibrinogen or macromolecular depletion | Low fibrinogen, abnormal coagulation tests, bleeding | Know the secondary-filter molecular cutoff and targets; assess baseline/post-treatment fibrinogen according to protocol | Re-evaluate treatment intensity and bleeding management before repeat procedures. |
| Membrane fouling / TMP rise | Pressure trend rise, low secondary-filtration yield | Check flow, membrane fouling and anticoagulation; respect approved TMP and pore-size selection | Investigate and stop/replace filter if unsafe; never bypass alarms. |
| Hemolysis / cell trauma | Dark plasma, abnormal labs, circuit pressure extremes | Follow both stage pressure profiles, device setup and filter compatibility | Assess urgently and stop harmful circuit. |
| Albumin loss / volume imbalance | Low oncotic tolerance, hemodynamic instability | Record waste volume and quantify replacement need using validated protocol | Review fluid/protein replacement with clinical team. |
| Reduced exposure to other circulating proteins or medicines | Unanticipated target or non-target loss | Review patient indication and membrane-specific sieving; plan medication timing and measured endpoints | Use tailored monitoring, not generic DFPP assumptions. |
12. Extracorporeal photopheresis (ECP)
ECP involves leukocyte collection, photoactivation with a psoralen/UVA system where applicable, and reinfusion. Closed and offline/multistep configurations have different validated components and anticoagulation workflows. [R18] [R19] [R20]
| Complication / mechanism | Warning signs | Prevention & monitoring | Immediate action / escalation |
|---|---|---|---|
| Hypotension / extracorporeal-volume intolerance | Dizziness, low BP, reduced consciousness | Assess weight, hematocrit, baseline perfusion and tolerated extracorporeal volume; monitor throughout collection/reinfusion | Pause the procedure and manage hemodynamic instability per local protocol. |
| Anemia or thrombocytopenia with repeat sessions | Progressive Hb or platelet decline, fatigue or bleeding | Plan CBC checks across treatment cycles; monitor cumulative blood loss | Reassess timing and any intervention with treating team. |
| Pyrexia or transient inflammatory symptoms after reinfusion | Post-procedure fever, chills or erythroderma | Provide post-treatment advice and institutional fever/infection assessment instructions | Evaluate other causes (including infection/transfusion reactions) rather than attributing every fever to ECP. |
| Psoralen sensitivity and UVA-related risk | Photosensitivity, severe history of psoralen allergy; device-label contraindications | Screen specific methoxsalen/psoralen contraindications and photosensitivity; verify product and dose under IFU; follow labeled eye/skin protection guidance | Stop suspected drug reaction and follow urgent allergy and ophthalmic/skin assessment pathways as indicated. |
| Anticoagulation or device-processing errors | Unexpected bleeding, clotting, bag/line misidentification or photoactivation failure | Use equipment-specific heparin/citrate method, identity checks and validated UVA exposure; avoid non-certified offline bags | Stop unvalidated cell processing; investigate deviations, maintain traceability and report. |
Before treatment
- Verify ECP platform, leukocyte handling/UV validation, psoralen contraindications, hematocrit and anticoagulation.
During treatment
- Monitor patient tolerance and integrity of closed/offline handling; confirm photoactivation workflow.
After treatment
- Monitor access and post-infusion reactions; follow photosensitivity precautions and course-level blood counts as indicated.
13. Monitoring plan — before, during & after
Document an individualized monitoring frequency and escalation owner. The table is a monitoring checklist; it does not impose universal laboratory thresholds or replace the IFU. [R2] [R3] [R16]
| Domain | Before treatment | During treatment | After / between sessions |
|---|---|---|---|
| Hemodynamics | Baseline BP/HR, volume and comorbidities | Vitals, symptoms, perfusion, balance; frequency per acuity | Volume status, delayed hypotension or syncope |
| Blood counts / coagulation | CBC, platelets, PT/INR, aPTT and fibrinogen by technique/risk | Visible bleeding, access site; tests if clinically needed | Trend Hb, platelets, fibrinogen and coagulation after protein-removing techniques |
| Electrolytes / citrate | Ionized Ca, Mg, K where citrate exposure possible | Symptoms, ECG if indicated, repeat iCa by validated citrate protocol | Evaluate delayed electrolyte effects in high-risk patients |
| Circuits and access | Connection trace, primes, air detectors, flows, kit/lot, access patency | Pressures/TMP, blood and plasma flow, alarms, clotting and hemolysis | Blood-return completeness, residual loss, access and device event log |
| Therapeutic proteins / drug exposure | Target antibodies/proteins, medications critical to treat | Possible non-target adsorption, infusion timing, target clinical response | Selected IgG/fibrinogen/albumin or drug levels as indicated |
| ECP-specific | CBC, photosensitivity/psoralen checks, approved system | Cell collection, UVA processing, return and fluid tolerance | Delayed fever, photosensitivity and repeated-session counts |
14. Response to complications — safety priorities
The following are safety escalation categories, not self-contained resuscitation protocols. Activate the institution’s emergency response pathways as clinically indicated. [R2] [R3] [R4]
| Presentation | Immediate priority | Do not overlook |
|---|---|---|
| Suspected anaphylaxis / airway compromise | Stop implicated exposure/procedure, activate emergency protocol, support airway/hemodynamics | Donor plasma, replacement products, resin/material reactions, medication triggers |
| Symptomatic severe citrate toxicity / arrhythmia | Urgent cardiovascular assessment and calcium/electrolyte protocol; reassess citrate delivery | Citrate in donor plasma as well as circuit; liver dysfunction and repeated exposure |
| Severe hypotension or bleeding | Interrupt/pause procedure, assess hemorrhage, allergy, shock and volume status; urgent clinician evaluation | Coagulation-factor depletion, heparin-related effects, catheter trauma |
| Possible hemolysis / massive circuit clot / air entry | Follow machine emergency stop and blood-return policy; assess patient and labs | Do not automatically reinfuse compromised circuit blood |
| Catheter thrombosis, infection or embolism | Stop unsafe use of access and obtain prompt medical assessment | Access-related complications may emerge after the session |
| Unexpected hemoglobin/platelet or protein loss | Review modality-related losses, underlying disease and drugs; compare baseline data | No single post-procedure laboratory threshold works for all devices |
| Suspected loss of therapeutic drug exposure | Contact treating prescriber/pharmacist, consider level monitoring where available | Critical antibiotics and other narrow-therapeutic-index drugs may be affected |
15. Safety checklist & documentation
Use this checklist with the exact product IFU and the clinical safety worksheet. Completing it does not authorize the procedure.
16. Safety & complications worksheet — complete & copy
Fill in a technique-specific risk and safety plan for team review. Fields are local to the current page: nothing is sent to a server or saved automatically. Review and sign the final clinical order in the authorized medical record.
17. Scientific guidelines, studies & source notes
Every technique chapter and matrix row links to its supporting references using the [R#] markers. A source can describe adverse-event patterns without validating a specific device protocol. The original article, current IFU and national labeling must be consulted directly.
- R1. American Society for Apheresis (ASFA), 2026. Guidelines on the Use of Therapeutic Apheresis in Clinical Practice — Tenth Special Issue. 2026 guideline; indication-specific evidence and modality definitions. Open source ↗
- R2. British Society for Haematology, Howell et al., 2025. Apheresis procedures for the treatment of patients and collection of cellular therapy products. Guideline; citrate, allergy, hypotension, access and coagulopathy prevention. Open source ↗
- R3. Canadian Blood Services. Therapeutic apheresis — Clinical Guide. Professional reference; adverse events, blood products and access-related complications. Open source ↗
- R4. Mokrzycki & Balogun, 2011. Therapeutic apheresis: a review of complications and recommendations for prevention and management. Clinical review; shared procedural hazards. Open source ↗
- R5. Therapeutic Plasma Exchange — A Practical Guide, 2026. Therapeutic Plasma Exchange — A Practical Guide. Clinical review; TPE planning, anticoagulation and adverse-event surveillance. Open source ↗
- R6. Ahmed & Kaplan, 2020. Therapeutic Plasma Exchange Using Membrane Plasma Separation. Technique review; membrane pressure, hemolysis, bradykinin and fluid/electrolyte risks. Open source ↗
- R7. Jouffroy et al., 2017. Plasma exchange in the intensive care unit: Technical aspects and complications. Observational comparison; fluid reactions and membrane-circuit clotting. Open source ↗
- R8. Tolerance of centrifugal TPE, 2021. Tolerance and complications of therapeutic plasma exchange by centrifugation: A single-center experience. Observational study; hypotension, hypocalcemia and plasma-related allergic events. Open source ↗
- R9. Gao et al., 2026. Risk factors for hypotension and technical complications during single-session DFPP: 1,022 procedures. Retrospective DFPP study; hypotension, TMP rise and equipment-related events. Open source ↗
- R10. Complications of double-filtration plasmapheresis, 2004. Complications of double-filtration plasmapheresis. Observational study; hemolysis, bleeding and access complications. Open source ↗
- R11. Recent Progress in Double-Filtration Plasmapheresis, 2025. Recent Progress in Double Filtration Plasmapheresis. Narrative review; secondary-filter selectivity, albumin/fibrinogen losses. Open source ↗
- R12. Therapeutic Plasmapheresis — Review, 2023. Therapeutic Plasmapheresis: A Revision of Literature. Modality comparison, safety, coagulation, ACE-inhibitor/bradykinin considerations. Open source ↗
- R13. Immunoadsorption clinical review, 2022. Performance, clinical effectiveness, and safety of immunoadsorption in a wide range of indications. Clinical study/review; IA tolerability and procedural risks. Open source ↗
- R14. Immunoadsorption sham-controlled trial, 2026. Immunoadsorption versus sham treatment for post-COVID syndrome: randomized crossover trial. Trial; IA events including central-catheter thromboses, circuit clotting and infections; not proof of efficacy for other diseases. Open source ↗
- R15. Lan et al., 2024. Safety profile of dual plasma molecular adsorption system in liver failure and hyperbilirubinemia. Retrospective DPMAS cohort; bleeding, hypotension and coagulation changes. Open source ↗
- R16. 30th Acute Disease Quality Initiative (ADQI), 2024. Hemoadsorption: consensus report. Expert consensus; off-target drug adsorption, cytopenias, circuit monitoring and evidence gaps. Open source ↗
- R17. EUPHRATES trial, 2018. Effect of targeted polymyxin B hemoperfusion on 28-day mortality in septic shock. Randomized trial; device-related events and clinical-efficacy limitations. Open source ↗
- R18. European Dermatology Forum, 2020. Updated guidelines on extracorporeal photopheresis, Part 1. Professional guideline; ECP tolerability, hematologic and photosensitivity cautions. Open source ↗
- R19. THERAKOS. CELLEX photopheresis system — Important Safety Information. Manufacturer safety/IFU-context statement; hypotension, pyrexia, contraindications. Open source ↗
- R20. U.S. FDA, UVADEX (methoxsalen) label. UVADEX sterile solution — prescribing information. Regulatory drug label; psoralen precautions, warnings and photopheresis adverse events. Open source ↗
- R21. Kosmadakis, 2022. Rheopheresis: A narrative review. Narrative review; cascade/rheofiltration targets and modality distinction. Open source ↗
- R22. EUPHRATES/TIGRIS evidence update, 2026. Polymyxin B haemoadsorption in endotoxic septic shock (TIGRIS): phase 3 randomized trial. Randomized study in a selected population; clinical-effectiveness context, not a device-wide safety rule. Open source ↗