Clinical reference · Therapeutic extracorporeal procedures

Complications & Safety in Therapeutic Apheresis & Adsorption

Technique-Specific Practical Safety Guide — Prevention, Early Detection, Monitoring & Escalation

All techniques are shown. Choose a treatment to focus the safety matrix, chapter and references.

1. Scope & evidence boundaries

Clinical-use boundary. Educational safety reference for trained therapeutic apheresis teams. Not a validated clinical order set, a device operating manual or an emergency algorithm. Local clinical approval and the current model-specific instructions for use (IFU) take precedence. Adapt for the patient, disease, age, circuit and regional regulations.

Covered: whole-blood extracorporeal circuit safety; therapeutic plasma exchange by centrifugation or membrane filtration; plasma adsorption; immunoadsorption; whole-blood hemoadsorption; double-filtration plasmapheresis (DFPP); cascade filtration/rheopheresis; and extracorporeal photopheresis (ECP). Not covered: hemodialysis or renal replacement anticoagulation protocols.

Terminology: “whole blood” below means safety of an extracorporeal blood circuit, not a stand-alone treatment indication. DFPP and cascade filtration commonly describe the same two-filter principle; they have separate navigation entries for practical usability, not because they require fundamentally different risk protocols. [R1] [R2] [R12]

Patient safety
Hemodynamic status, bleeding risk, anticoagulant reactions, allergy and access injury
Circuit safety
Validated blood path, pressure limits, air detection, hemolysis and clot formation
Therapy-specific safety
Plasma protein loss, sorbent interactions, drug removal, cell depletion or methoxsalen exposure
Evidence hierarchy: current IFU/regulatory labeling → professional society guidance → technique-specific trials and observational studies → mechanistic reviews. A trial of one cartridge is not a universal safety rate for all other devices. [R1] [R2] [R16]

2. Technique-specific safety matrix

Choose one technique using the selector above. The matrix, navigation and detailed safety chapter will change together. Shared triage and worksheet sections remain visible for every choice.

TechniquePrincipal complicationsPriority safeguardsSupporting evidence
Whole-blood extracorporeal circuitAir/clot/hemolysis; access injury; unexpected blood lossTrace and prime; pressure/air alarm checks; trained return-of-blood procedure[R2] [R4]
TPE — centrifugationCitrate hypocalcemia; plasma allergy; hypotension; coagulation-factor depletionIonized Ca and symptoms; blood-product checks; volume and fibrinogen plan[R2] [R8]
TPE — membrane filtrationTransmembrane pressure rise; filter clotting/hemolysis; hypotension; coagulopathyMonitor TMP/hemolysis; approved circuit anticoagulation; fluid/coagulation plan[R6] [R7]
Plasma adsorption / DPMASBleeding or coagulopathy; hypotension; allergic reaction; cartridge clottingPlatelets/fibrinogen/INR; pressures; access and fluid monitoring[R12] [R15]
ImmunoadsorptionAdsorber/clotting or allergy; antibody/protein depletion; catheter eventsCheck ligand and regeneration IFU; Ig targets; access and infection surveillance[R12] [R13] [R14]
Whole-blood hemoadsorptionThrombocytopenia; drug removal; pressure/clotting; hypersensitivityCBC and medication review; pressures; validated cartridge/anticoagulation[R16] [R17]
DFPPSecondary-filter pressure rise; hemolysis; fibrinogen/Ig loss; hypotensionTMP1/TMP2 trend; coagulation/fibrinogen; albumin and balance assessment[R9] [R10]
Cascade filtration / rheopheresisDFPP-type circuit risks; target-dependent protein depletionVerify secondary-filter pore selectivity, pressure and replacement plan[R11] [R21]
Extracorporeal photopheresisHypotension; cytopenia; fever; drug/light sensitivityExtracorporeal volume; CBC; psoralen screening; platform IFU[R18] [R19] [R20]
All hazards are possible complications, not a prediction that the event will occur. Absolute rates are device-, indication- and population-dependent; use local adverse-event reporting rather than extrapolating frequencies from a different device. [R2] [R4]

3. Pre-procedure clinical & technical safety screen

Before treatment

  • Confirm indication, rationale, expected benefit and consent; identify special vulnerability (hemodynamic instability, allergy, bleeding, pregnancy and low body weight where relevant).
  • Review CBC/platelets, PT/INR, aPTT, fibrinogen, electrolytes (including ionized Ca and Mg when citrate may be used), renal/liver function where clinically relevant.
  • Check allergy to albumin/plasma, anticoagulants, membrane/sorbent and psoralen where relevant; assess previous HIT and ACE-inhibitor exposure.
  • Choose access and confirm catheter position/patency, asepsis, compatible machine/kit/lot and approved anticoagulation strategy.
  • Document fluid/replacement, medication timing, blood conservation and contingency plans.

During treatment

  • Baseline and interval vitals, symptoms, fluid balance, access and circuit pressure/air alarms; observe the patient—not alarms alone.
  • Check for citrate symptoms, bleeding, anaphylaxis, clotting, hemolysis, and target-specific losses.
  • Escalate abnormal trends according to the local protocol; obtain urgent laboratory tests when clinically indicated.

After treatment

  • Recheck clinical state, access site, fluid balance and indicated CBC/coagulation/electrolytes.
  • Document actual treatment delivered, product exposure, adverse events, medication timing and a plan for subsequent sessions.
  • Report serious or device-related events to the institutional safety team and manufacturer as appropriate.

A system-wide safety check cannot replace the technique-specific IFU. [R2] [R3] [R4]

4. Whole-blood extracorporeal circuit — common hazards

Applies to any procedure that withdraws and returns whole blood through tubing, pumps and vascular access. This is a cross-cutting circuit safety category, not a separate blood purification treatment. [R2] [R4]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Vascular access bleeding, hematoma, thrombosis or infectionOozing, swelling, difficult aspiration, pain, erythema, feverAseptic access, secure dressings, verify patency and anticoagulation plan; reassess access each sessionPause or interrupt as appropriate; assess bleeding/thrombosis/infection; escalate urgently if major event.
Air entry / embolismAir alarm, air in tubing, abrupt dyspnea, chest symptoms or collapseValidated air detectors, intact connections, correct priming and alarm checksStop per emergency circuit protocol; initiate immediate clinical emergency response; return of circuit blood depends on event and IFU.
Clotting / rising circuit pressureVisible clots; pressure increases; flow failureCorrect anticoagulation, air removal, pressure monitoring and allowed flow configurationDo not bypass alarms; stop or replace circuit per IFU. Do not return potentially compromised circuit blood without clinical/device authorization.
Hemolysis / mechanical blood traumaDark plasma, red/brown effluent, pain, unexplained K rise or instabilityCheck tubing kinks, inappropriate pressures and machine/pump setupUrgently halt suspected causative circuit and assess hemolysis, electrolytes and clinical stability.
Extracorporeal blood loss, hypovolemia, hypothermiaHypotension, tachycardia, cold extremities, low return volumeDocument extracorporeal volume and returned blood; warming/fluid plan where appropriateAssess and support perfusion, temperature and fluid balance; decide continuation after reassessment.
Never assume that trapped/clotted, hemolyzed or potentially contaminated blood can be safely returned. Follow the exact device emergency procedure. [R2] [R3] [R4]

5. TPE by centrifugation — complications & prevention

Centrifugal therapeutic plasma exchange commonly uses citrate at the blood inlet; reaction patterns are strongly influenced by replacement fluid and total citrate exposure. [R2] [R5] [R8]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Citrate-associated hypocalcemia ± hypomagnesemiaPerioral tingling, cramps, nausea; severe tetany, QT prolongation or arrhythmiaScreen risk factors; monitor symptoms and ionized Ca (plus Mg where indicated); validate citrate/calcium regimen and source of citrate in donor plasmaFollow local citrate-toxicity pathway: reassess procedure/citrate exposure and treat symptomatic electrolyte disturbance urgently.
Hypotension or vasovagal reactionDizziness, sweating, low BP, nausea or reduced consciousnessHemodynamic risk screen; appropriate replacement balance and extracorporeal-volume planningPause procedure and assess volume status, allergic causes, arrhythmia and other shock causes; escalate.
Allergy / anaphylaxis / transfusion reactionRash, wheeze, flushing, fever or collapse after plasma exposureVerify identity and plasma product; screen allergy history; have reaction protocol and resuscitation equipmentStop exposure and activate institutional transfusion/anaphylaxis response; preserve product data for investigation.
Coagulopathy and loss of protective plasma constituentsNew bleeding, fibrinogen fall, prolonged coagulation testingBaseline and repeat fibrinogen/platelets/coagulation tests in high-risk or repeated sessions; individualize replacement strategyReassess further exchanges and blood-product needs with treating physician.
Medication removal or interactionsLower-than-expected drug effect or level after exchangeReview protein binding, distribution volume and dose timing of critical medicationsUse therapeutic drug monitoring or specialist pharmacist plan when indicated.

Before treatment

  • Confirm citrate and replacement-fluid plan; allergy, coagulation and calcium baseline.

During treatment

  • Assess neurologic/cardiac symptoms and vitals; check iCa at locally required intervals; observe for transfusion reactions.

After treatment

  • Examine puncture sites; reassess fibrinogen/electrolytes when indicated; review drug timing and next treatment.
Evidence example: a retrospective centrifugal-TPE cohort reported hypotension, hypocalcemia and allergic reactions, with rates dependent on replacement fluid. Do not interpret its event rates as universal machine rates. [R8]

6. TPE by membrane filtration — complications & prevention

Primary membrane plasma separation has particular concerns about transmembrane pressure, progressive filter clotting and mechanical hemolysis. Some filters may increase bradykinin-related risk in patients exposed to ACE inhibitors. [R6] [R7] [R12]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Excess transmembrane pressure / filter clottingTMP increase, falling plasma flow, alarm or visible clotUse approved plasma fraction and blood/plasma flow ranges; correct anticoagulation and pressure trend monitoringDo not increase pressure past IFU limits. Investigate and replace circuit/filter when indicated.
HemolysisHemolyzed effluent, dark plasma, unexpected pain, instability or electrolyte changeAvoid excessive TMP, pressure gradients and tubing obstruction; monitor color and circuit dataStop suspected harmful circuit; promptly evaluate hemolysis and cardiac/electrolyte risk.
Hypotension / fluid imbalanceBP fall, poor perfusion, tachycardiaPlan replacement/fluid balance, blood-flow tolerance and access compatibilityAssess cause and restore hemodynamic stability per institutional protocol.
Coagulation-factor/fibrinogen depletionBleeding or low fibrinogen after plasma removalFollow CBC, fibrinogen and coagulation tests based on patient risk and repeat exchangesReassess replacement and continuation before further exchange.
Bradykinin-associated reaction (selected membranes / ACE inhibitors)Abrupt flushing, hypotension, dyspneaReview ACE inhibitor exposure and the exact membrane IFU; follow any required drug interruption windowTreat as a severe reaction until proven otherwise; stop the implicated system and escalate.
Replacement-plasma reactions or citrate exposureAllergy, respiratory symptoms or tingling during plasma replacementProduct verification; citrate and calcium awareness even if the circuit uses heparinFollow transfusion-reaction and electrolyte evaluation pathways.
A retrospective ICU series reported filter clotting in a subset of membrane procedures; this is an indication to trend pressures, not to apply an unsupported standard anticoagulation dose. [R7]

7. Plasma adsorption / bilirubin adsorption / DPMAS

In plasma adsorption, separated plasma passes through one or more sorbent columns and returns to the patient. Risk depends on the ligand/resin, cartridge capacity, anticoagulant and the patient’s underlying coagulopathy; DPMAS liver-failure populations need special vigilance. [R12] [R15]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Bleeding / worse coagulationPuncture bleeding, mucosal bleed, falling fibrinogen or platelets, rising INRDocument baseline CBC, coagulation/fibrinogen and repeat with clinical severity; assess adsorption-related and liver-related lossesReassess anticoagulation and procedural continuation; manage bleeding under local protocol.
Hypotension / extracorporeal-volume intoleranceEarly BP drop, tachycardia, dizzinessPre-treatment hemodynamic assessment, appropriate priming and careful fluid planPause and assess perfusion, volume, allergic and bleeding causes.
Allergy / biocompatibility reactionFlushing, urticaria, bronchospasmConfirm resin/ligand history, priming/wash and approved materialsTerminate suspected triggering exposure and treat reaction urgently if severe.
Column clotting / pressure riseInlet/outlet pressure changes, loss of plasma flowValidated anticoagulant and cartridge orientation; trend pressures, respect cartridge lifespanStop or replace malfunctioning column per IFU; do not bypass alarms.
Unintended protein or medication adsorptionUnexpected laboratory/protein decline or loss of drug effectReview target and potential off-target binding; monitor clinically critical proteins/drug concentrations where feasibleConsider dose timing or adjustment with pharmacy/clinical team; do not assume selective adsorption is perfectly specific.
DPMAS evidence: one retrospective cohort reported bleeding, hypotension, allergic reactions and infection; the findings are not direct risk estimates for other plasma adsorbers or healthy patients. [R15]

Before treatment

  • Name the exact sorbents and target, separator, coagulation status, anticoagulation and emergency limits.

During treatment

  • Monitor access, separated-plasma flow, column pressures, hemodynamics and bleeding.

After treatment

  • Document delivered plasma volume, pressure events, blood loss, CBC/coagulation changes and medication review.

8. Immunoadsorption (including ABO-specific)

Immunoadsorption removes selected antibodies/immune complexes after plasma separation. Single-use and regenerable columns have different safety steps. ABO-specific columns also require target-antibody monitoring, and the overall safety assessment depends on the transplant or disease context. [R1] [R13] [R14]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Bradykinin-related severe hypotension in at-risk circuitsSudden flushing, dyspnea, hypotension, bradycardiaReview ACE inhibitors and the exact adsorption/filter IFU; apply manufacturer-defined restrictionsStop the relevant exposure and manage anaphylactoid/shock presentation immediately.
Column-related hypersensitivityUrticaria, airway symptoms, hypotensionConfirm ligand/material history, proper priming and approved column combinationsManage reaction and report potential device issue; do not restart without review.
Excess antibody/protein depletion or immunologic vulnerabilityMarked Ig reduction, infection or poor vaccine/Ig therapy timingTrack target antibodies/IgG (and other Ig classes if relevant); coordinate IVIG/biologics; assess infection riskAdjust session plan and adjunct therapy with immunology/transplant service.
Regeneration/rinse or column-switching failureUnexpected alarm, conductivity/pH change, wrong valve stateUse only validated regeneration sequence and leak/rinse safeguards; two-person verification when requiredStop device according to IFU; do not allow regeneration solution or unverified path to reach patient.
Access-related thrombosis and circuit clottingPoor flow, catheter swelling/pain, access alarms, visible clotAccess assessment and anticoagulation per platform; review symptoms after multiple sessionsEvaluate suspected thrombosis and access complications promptly.
A sham-controlled IA trial reported catheter-associated jugular thromboses and some circuit clotting; the setting and patient selection differ from standard transplant IA. [R14]

Before treatment

  • Identify ligand/ABO specificity, planned antibody monitoring, ACE-inhibitor restrictions, regeneration and rinse checks.

During treatment

  • Watch for abrupt hypotension, column alarms, clotting and adverse reactions.

After treatment

  • Reassess target titers and clinically relevant Ig/proteins, catheter sites and timing of IVIG/immunosuppression.

9. Whole-blood hemoadsorption / hemoperfusion

A whole-blood sorbent cartridge is perfused directly with blood. Unlike plasma adsorption, no plasma separator is necessarily present. This chapter applies to hemoadsorption circuits without introducing renal replacement treatment protocols. Cartridge chemistry, indication and integration materially affect safety. [R16] [R17]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Thrombocytopenia, anemia, leukocyte/protein lossPlatelet/Hb/WBC decline, bleeding or repeated-session blood lossCheck CBC before and after sessions as clinically indicated; document blood losses and cartridge biocompatibilityInvestigate cause including underlying disease, anticoagulants and sorbent effects; reassess subsequent treatment.
Unintended adsorption of medicinesDrug level falls or loss of effect, e.g. time-critical antibiotic exposurePharmacist review of antibiotics, antiepileptics, anticoagulants and narrow-therapeutic-index drugs; use drug levels if availableOptimize dosing/timing with treating team; avoid assumptions about adsorption for a specific drug or cartridge.
Cartridge clotting / excessive resistanceRising pressure gradient, flow alarms, visible clotConfirm IFU-allowed cartridge location, priming, blood flow and anticoagulation; follow pressure limitsStop/reconfigure only under approved procedure; do not ignore pressure alarms.
Hypotension or systemic reactionBP drop, flushing, dyspnea, escalating support needsAssess hemodynamics, volume status and compatibility; maintain emergency response readinessPause treatment, assess shock/anaphylaxis and activate urgent escalation if needed.
Blood trauma, air embolism and vascular injuryDiscoloration, air alarms, catheter problemsUse whole-blood circuit checks and trained operators; trace lines and air detectorsEmergency circuit response according to IFU; evaluate clinical consequences.
Clinical effectiveness is indication- and cartridge-specific. The EUPHRATES RCT did not show 28-day mortality benefit for polymyxin-B hemoperfusion in its overall population, while later trials studied narrower populations. These outcome studies do not justify routine hemoadsorption for all patients or translate into universal complication rates. [R17] [R22]

Before treatment

  • Exact cartridge/indication, bleeding and cytopenia baseline; pharmacist review of essential medications.

During treatment

  • Trend pressure gradient, flow, clinical stability and device time limits; assess clotting and new reactions.

After treatment

  • CBC and clinically relevant drug levels; check access and medication exposure; report device-related reactions.

10. Double-filtration plasmapheresis (DFPP)

DFPP passes separated plasma through a secondary fractionation membrane to remove larger molecules; the retained fraction is discarded and the filtrate returned. The filter is not completely selective: fibrinogen, immunoglobulins and some albumin may be lost. [R9] [R10] [R11]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Secondary-filter TMP rise / occlusionTMP2 rise, reduced permeate flow, pressure alarmsVerify secondary-filter choice, blood/plasma rates, anticoagulation and accepted pressure limits; trend both filtration stagesDo not force pressure/flow beyond IFU; investigate clotting and exchange filters as indicated.
HemolysisDark plasma, hemolysis alarm, sudden electrolyte changeMonitor excessive TMP or shear, tubing integrity and filter compatibilityStop suspected traumatic circuit; evaluate blood trauma and clinical impact.
Hypofibrinogenemia / coagulation-factor depletionUnexpected fibrinogen drop, bleeding or prolonged coagulation testsUse indication-specific pretreatment and repeat fibrinogen/coagulation checks; plan replacement/blood products individuallyReconsider further filtration and treat clinically meaningful factor loss per local protocol.
Albumin/Ig losses and fluid imbalanceLow albumin, edema, hypo-oncotic symptoms or increased infection susceptibility after repeated sessionsQuantify discarded fraction and replacement need; check proteins and antibody levels when relevantAdjust treatment schedule, fluids and planned adjunct therapies.
Hypotension / access failureBP fall, nausea, unstable flowRisk-adapt extracorporeal volume and access; monitor continuouslyAssess hypoperfusion and vascular access; interrupt as clinically appropriate.
A 2026 retrospective study of 1,022 DFPP procedures identified hypotension and technical complications, including rising TMP2; a separate earlier study reported hemolysis. Rates differ by equipment, era and clinical population. [R9] [R10]

11. Cascade filtration / rheopheresis — specific safeguards

Cascade filtration is often a synonym for DFPP; rheopheresis denotes a filter/target-specific form used to remove selected high-molecular-weight substances. It is listed separately here so users can select “cascade filtration”, but the fundamental pressure, hemolysis, access and anticoagulation precautions overlap with DFPP. [R11] [R12] [R21]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Unintended fibrinogen or macromolecular depletionLow fibrinogen, abnormal coagulation tests, bleedingKnow the secondary-filter molecular cutoff and targets; assess baseline/post-treatment fibrinogen according to protocolRe-evaluate treatment intensity and bleeding management before repeat procedures.
Membrane fouling / TMP risePressure trend rise, low secondary-filtration yieldCheck flow, membrane fouling and anticoagulation; respect approved TMP and pore-size selectionInvestigate and stop/replace filter if unsafe; never bypass alarms.
Hemolysis / cell traumaDark plasma, abnormal labs, circuit pressure extremesFollow both stage pressure profiles, device setup and filter compatibilityAssess urgently and stop harmful circuit.
Albumin loss / volume imbalanceLow oncotic tolerance, hemodynamic instabilityRecord waste volume and quantify replacement need using validated protocolReview fluid/protein replacement with clinical team.
Reduced exposure to other circulating proteins or medicinesUnanticipated target or non-target lossReview patient indication and membrane-specific sieving; plan medication timing and measured endpointsUse tailored monitoring, not generic DFPP assumptions.
Do not treat “cascade” as an alternative anticoagulation standard. It is a secondary-filtration configuration. All flows, pressure thresholds and anticoagulation are device-specific. [R11] [R21]

12. Extracorporeal photopheresis (ECP)

ECP involves leukocyte collection, photoactivation with a psoralen/UVA system where applicable, and reinfusion. Closed and offline/multistep configurations have different validated components and anticoagulation workflows. [R18] [R19] [R20]

Complication / mechanismWarning signsPrevention & monitoringImmediate action / escalation
Hypotension / extracorporeal-volume intoleranceDizziness, low BP, reduced consciousnessAssess weight, hematocrit, baseline perfusion and tolerated extracorporeal volume; monitor throughout collection/reinfusionPause the procedure and manage hemodynamic instability per local protocol.
Anemia or thrombocytopenia with repeat sessionsProgressive Hb or platelet decline, fatigue or bleedingPlan CBC checks across treatment cycles; monitor cumulative blood lossReassess timing and any intervention with treating team.
Pyrexia or transient inflammatory symptoms after reinfusionPost-procedure fever, chills or erythrodermaProvide post-treatment advice and institutional fever/infection assessment instructionsEvaluate other causes (including infection/transfusion reactions) rather than attributing every fever to ECP.
Psoralen sensitivity and UVA-related riskPhotosensitivity, severe history of psoralen allergy; device-label contraindicationsScreen specific methoxsalen/psoralen contraindications and photosensitivity; verify product and dose under IFU; follow labeled eye/skin protection guidanceStop suspected drug reaction and follow urgent allergy and ophthalmic/skin assessment pathways as indicated.
Anticoagulation or device-processing errorsUnexpected bleeding, clotting, bag/line misidentification or photoactivation failureUse equipment-specific heparin/citrate method, identity checks and validated UVA exposure; avoid non-certified offline bagsStop unvalidated cell processing; investigate deviations, maintain traceability and report.
Some manufacturer contraindications (including psoralen sensitivity and inability to tolerate extracorporeal-volume changes) are platform- and labeling-specific. Check the current THERAKOS CELLEX/UVADEX labeling or the actual system in use. [R18] [R19] [R20]

Before treatment

  • Verify ECP platform, leukocyte handling/UV validation, psoralen contraindications, hematocrit and anticoagulation.

During treatment

  • Monitor patient tolerance and integrity of closed/offline handling; confirm photoactivation workflow.

After treatment

  • Monitor access and post-infusion reactions; follow photosensitivity precautions and course-level blood counts as indicated.

13. Monitoring plan — before, during & after

Document an individualized monitoring frequency and escalation owner. The table is a monitoring checklist; it does not impose universal laboratory thresholds or replace the IFU. [R2] [R3] [R16]

DomainBefore treatmentDuring treatmentAfter / between sessions
HemodynamicsBaseline BP/HR, volume and comorbiditiesVitals, symptoms, perfusion, balance; frequency per acuityVolume status, delayed hypotension or syncope
Blood counts / coagulationCBC, platelets, PT/INR, aPTT and fibrinogen by technique/riskVisible bleeding, access site; tests if clinically neededTrend Hb, platelets, fibrinogen and coagulation after protein-removing techniques
Electrolytes / citrateIonized Ca, Mg, K where citrate exposure possibleSymptoms, ECG if indicated, repeat iCa by validated citrate protocolEvaluate delayed electrolyte effects in high-risk patients
Circuits and accessConnection trace, primes, air detectors, flows, kit/lot, access patencyPressures/TMP, blood and plasma flow, alarms, clotting and hemolysisBlood-return completeness, residual loss, access and device event log
Therapeutic proteins / drug exposureTarget antibodies/proteins, medications critical to treatPossible non-target adsorption, infusion timing, target clinical responseSelected IgG/fibrinogen/albumin or drug levels as indicated
ECP-specificCBC, photosensitivity/psoralen checks, approved systemCell collection, UVA processing, return and fluid toleranceDelayed fever, photosensitivity and repeated-session counts
Not every laboratory test is needed in every patient. The treating clinical team sets tests and frequencies according to modality, IFU, baseline condition and symptoms. [R2] [R5]

14. Response to complications — safety priorities

The following are safety escalation categories, not self-contained resuscitation protocols. Activate the institution’s emergency response pathways as clinically indicated. [R2] [R3] [R4]

PresentationImmediate priorityDo not overlook
Suspected anaphylaxis / airway compromiseStop implicated exposure/procedure, activate emergency protocol, support airway/hemodynamicsDonor plasma, replacement products, resin/material reactions, medication triggers
Symptomatic severe citrate toxicity / arrhythmiaUrgent cardiovascular assessment and calcium/electrolyte protocol; reassess citrate deliveryCitrate in donor plasma as well as circuit; liver dysfunction and repeated exposure
Severe hypotension or bleedingInterrupt/pause procedure, assess hemorrhage, allergy, shock and volume status; urgent clinician evaluationCoagulation-factor depletion, heparin-related effects, catheter trauma
Possible hemolysis / massive circuit clot / air entryFollow machine emergency stop and blood-return policy; assess patient and labsDo not automatically reinfuse compromised circuit blood
Catheter thrombosis, infection or embolismStop unsafe use of access and obtain prompt medical assessmentAccess-related complications may emerge after the session
Unexpected hemoglobin/platelet or protein lossReview modality-related losses, underlying disease and drugs; compare baseline dataNo single post-procedure laboratory threshold works for all devices
Suspected loss of therapeutic drug exposureContact treating prescriber/pharmacist, consider level monitoring where availableCritical antibiotics and other narrow-therapeutic-index drugs may be affected
Return of blood from an interrupted circuit: this is a specific clinical/device decision. In suspected hemolysis, contamination, air ingress or clotting, do not apply a generic reinfusion rule. Follow current machine IFU and emergency protocol.

15. Safety checklist & documentation

Use this checklist with the exact product IFU and the clinical safety worksheet. Completing it does not authorize the procedure.

The checklist is a learning aid only; it does not certify treatment or replace two-person verification and institutional documentation policies. [R2] [R4]

16. Safety & complications worksheet — complete & copy

Fill in a technique-specific risk and safety plan for team review. Fields are local to the current page: nothing is sent to a server or saved automatically. Review and sign the final clinical order in the authorized medical record.

This template contains no preset orders, thresholds or doses. It is not a replacement for written institutional clinical documentation.

17. Scientific guidelines, studies & source notes

Every technique chapter and matrix row links to its supporting references using the [R#] markers. A source can describe adverse-event patterns without validating a specific device protocol. The original article, current IFU and national labeling must be consulted directly.

  1. R1. American Society for Apheresis (ASFA), 2026. Guidelines on the Use of Therapeutic Apheresis in Clinical Practice — Tenth Special Issue. 2026 guideline; indication-specific evidence and modality definitions. Open source ↗
  2. R2. British Society for Haematology, Howell et al., 2025. Apheresis procedures for the treatment of patients and collection of cellular therapy products. Guideline; citrate, allergy, hypotension, access and coagulopathy prevention. Open source ↗
  3. R3. Canadian Blood Services. Therapeutic apheresis — Clinical Guide. Professional reference; adverse events, blood products and access-related complications. Open source ↗
  4. R4. Mokrzycki & Balogun, 2011. Therapeutic apheresis: a review of complications and recommendations for prevention and management. Clinical review; shared procedural hazards. Open source ↗
  5. R5. Therapeutic Plasma Exchange — A Practical Guide, 2026. Therapeutic Plasma Exchange — A Practical Guide. Clinical review; TPE planning, anticoagulation and adverse-event surveillance. Open source ↗
  6. R6. Ahmed & Kaplan, 2020. Therapeutic Plasma Exchange Using Membrane Plasma Separation. Technique review; membrane pressure, hemolysis, bradykinin and fluid/electrolyte risks. Open source ↗
  7. R7. Jouffroy et al., 2017. Plasma exchange in the intensive care unit: Technical aspects and complications. Observational comparison; fluid reactions and membrane-circuit clotting. Open source ↗
  8. R8. Tolerance of centrifugal TPE, 2021. Tolerance and complications of therapeutic plasma exchange by centrifugation: A single-center experience. Observational study; hypotension, hypocalcemia and plasma-related allergic events. Open source ↗
  9. R9. Gao et al., 2026. Risk factors for hypotension and technical complications during single-session DFPP: 1,022 procedures. Retrospective DFPP study; hypotension, TMP rise and equipment-related events. Open source ↗
  10. R10. Complications of double-filtration plasmapheresis, 2004. Complications of double-filtration plasmapheresis. Observational study; hemolysis, bleeding and access complications. Open source ↗
  11. R11. Recent Progress in Double-Filtration Plasmapheresis, 2025. Recent Progress in Double Filtration Plasmapheresis. Narrative review; secondary-filter selectivity, albumin/fibrinogen losses. Open source ↗
  12. R12. Therapeutic Plasmapheresis — Review, 2023. Therapeutic Plasmapheresis: A Revision of Literature. Modality comparison, safety, coagulation, ACE-inhibitor/bradykinin considerations. Open source ↗
  13. R13. Immunoadsorption clinical review, 2022. Performance, clinical effectiveness, and safety of immunoadsorption in a wide range of indications. Clinical study/review; IA tolerability and procedural risks. Open source ↗
  14. R14. Immunoadsorption sham-controlled trial, 2026. Immunoadsorption versus sham treatment for post-COVID syndrome: randomized crossover trial. Trial; IA events including central-catheter thromboses, circuit clotting and infections; not proof of efficacy for other diseases. Open source ↗
  15. R15. Lan et al., 2024. Safety profile of dual plasma molecular adsorption system in liver failure and hyperbilirubinemia. Retrospective DPMAS cohort; bleeding, hypotension and coagulation changes. Open source ↗
  16. R16. 30th Acute Disease Quality Initiative (ADQI), 2024. Hemoadsorption: consensus report. Expert consensus; off-target drug adsorption, cytopenias, circuit monitoring and evidence gaps. Open source ↗
  17. R17. EUPHRATES trial, 2018. Effect of targeted polymyxin B hemoperfusion on 28-day mortality in septic shock. Randomized trial; device-related events and clinical-efficacy limitations. Open source ↗
  18. R18. European Dermatology Forum, 2020. Updated guidelines on extracorporeal photopheresis, Part 1. Professional guideline; ECP tolerability, hematologic and photosensitivity cautions. Open source ↗
  19. R19. THERAKOS. CELLEX photopheresis system — Important Safety Information. Manufacturer safety/IFU-context statement; hypotension, pyrexia, contraindications. Open source ↗
  20. R20. U.S. FDA, UVADEX (methoxsalen) label. UVADEX sterile solution — prescribing information. Regulatory drug label; psoralen precautions, warnings and photopheresis adverse events. Open source ↗
  21. R21. Kosmadakis, 2022. Rheopheresis: A narrative review. Narrative review; cascade/rheofiltration targets and modality distinction. Open source ↗
  22. R22. EUPHRATES/TIGRIS evidence update, 2026. Polymyxin B haemoadsorption in endotoxic septic shock (TIGRIS): phase 3 randomized trial. Randomized study in a selected population; clinical-effectiveness context, not a device-wide safety rule. Open source ↗
Sources include professional guidelines (ASFA 2026, BSH 2025), clinical trials, observational cohorts, reviews and manufacturer/FDA labeling. These categories have different evidentiary strength. The links need internet access; the guide itself is self-contained.