Clinical reference · Adult extracorporeal therapies

Anticoagulation in Therapeutic Apheresis & Adsorption

Practical Prescription Guide — Heparin, Citrate & Technique-Specific Protocols

Educational guide · Validate every dose, formulation and circuit against the current device IFU and local clinical protocol

Showing all techniques. Shared heparin, citrate, safety and worksheet sections remain available.

1. Scope, terminology & safety

Clinical-use boundary. This is a referenced educational prescribing framework for trained extracorporeal-therapy teams, not a validated bedside order set. Before use, reconcile the current country-specific IFU for the exact machine, cartridge, filter, anticoagulant solution and circuit with a hospital-approved protocol. Published example settings are not transferable between systems. This guide addresses adult practice; pediatric and obstetric care need separate protocols.

Anticoagulation is a circuit decision, not simply a disease decision. Determine whether blood is centrifuged, filtered through a membrane, passed through a plasma sorbent, or perfused through a whole-blood cartridge. Heparin and citrate have distinct systemic and regional effects. A patient already receiving systemic anticoagulation may need a different extracorporeal strategy. [R1] [R5]

UFH — unfractionated heparinSystemic anticoagulant. Monitor bleeding, platelets/HIT, aPTT/ACT or calibrated anti-Xa as applicable.
ACD-A — citrate in centrifuge circuitsAnticoagulant mixed with incoming blood. Monitor citrate symptoms and systemic ionized calcium.
RCA — regional citrate anticoagulationCitrate plus a coordinated calcium/solution plan. Post-filter calcium targets are specific to validated RCA protocols.
Critical exceptions: LIPOSORBER LA-15 labeling explicitly prohibits ACD citrate; CytoSorb stand-alone therapy requires heparin according to the manufacturer's FAQ, whereas citrate may be used in an appropriately configured combined circuit. [R9] [R10] [R7]
Scope: centrifugal and membrane therapeutic plasma exchange, DFPP/cascade filtration, plasma adsorption, whole-blood hemoadsorption, immunoadsorption and extracorporeal photopheresis. Renal replacement and hemodialysis protocols are outside this guide.

2. Technique-by-technique anticoagulation matrix

“Possible” means reported or used in validated configurations; it does not imply approval for every machine, filter, country or indication.

Technique / circuitHeparin (UFH)CitrateProtocol decision & evidence
Centrifugal TPEAlternative or adjunct in selected patientsCommon default: ACD-AUse separator-approved blood:ACD ratio; consider plasma replacement citrate load and calcium. [R1] [R5]
Membrane TPECommon optionPossible only with validated membrane/RCA setupCheck filter, replacement fluid, bleeding risk, citrate-calcium balance. [R1] [R5]
DFPP / cascade filtrationCommon comparatorRCA studied; not universalTwo filters, pressure/clotting and fibrinogen loss require separate checks. [R4]
Immunoadsorption (IA), incl. ABOAlone or in approved combinationOften ACD-A at centrifugal separationSeparator and adsorber must both permit chosen anticoagulant; a published ABO series used combined therapy. [R1] [R16]
Plasma adsorption: bilirubin, liver support, DPMASCommonly describedOnly if complete circuit validatedPlasma separator + adsorber + return line; no cross-brand universal dose. [R1] [R5]
LDL / Lp(a) plasma adsorptionDepends on device; required in LA-15 US systemProhibited in LA-15 US system; other systems differCheck exact lipoprotein device; do not generalize. [R9] [R10] [R19]
MONET / plasma lipid filtrationDevice-dependentDevice-dependentPlasma filter circuit; no generic dose or target extrapolated from DFPP. [R19]
H.E.L.P. LDL precipitationProcess-specific heparin and system-specific anticoagulationDo not substitute without validated IFUHeparin is part of the precipitation chemistry; not merely an optional anticoagulant. [R11] [R20]
Whole-blood hemoadsorption — stand-aloneCommon; device-specificNot automatically permittedCytoSorb stand-alone: manufacturer says heparin only. Other cartridges have different IFUs. [R7] [R17]
Hemoadsorption with ECMO / CPBOften governed by primary ECMO/CPB anticoagulation protocolNot a drop-in replacement for systemic ECMO anticoagulationDedicated ECMO/CPB team sets systemic targets; sorbent IFU must allow placement. [R7] [R8]
ECP (photopheresis)Some integrated systemsACD-A in many off-line circuits; option in current CELLEX EU materialsModel/software-specific. Do not transplant historical CELLEX heparin doses to other ECP systems. [R12] [R13] [R14] [R15]

3. Before starting: prescription & risk screen

  1. Confirm clinical indication and whether the proposed extracorporeal treatment itself is justified; anticoagulation does not establish treatment efficacy. For adult sepsis/septic shock, the 2026 Surviving Sepsis Campaign suggests against routine blood purification and polymyxin-B hemoperfusion. [R18]
  2. Record patient weight, active bleeding, surgery/procedures, history of HIT, prior citrate intolerance, shock/hypoperfusion, severe hepatic dysfunction, systemic anticoagulants and antiplatelets.
  3. Record baseline CBC/platelets, PT/INR, aPTT (and anti-Xa if indicated), fibrinogen for plasma-removing therapies, systemic ionized calcium, magnesium, sodium, bicarbonate/ABG and lactate if RCA considered. [R1] [R5]
  4. Document exact device model, software, circuit and disposable IFU revision, access, priming, blood/plasma flows, fluid composition, replacement and circuit volume.
  5. Choose anticoagulation: ACD-A, UFH, RCA, anticoagulant-free if appropriate and validated, or specialist non-heparin strategy for HIT. Document reason and monitoring plan. [R1]
  6. Assign clinician to approve start, monitor abnormalities, adjust infusion, and stop therapy if safety thresholds are breached.

4. UFH: prescription and monitoring protocol

Order elementWhat to prescribeSafety / evidence
PreparationUFH concentration (IU/mL), line and priming instructions, separate systemic infusion vs circuit injectionNever infer priming concentration from a different cartridge. [R1] [R13]
Initial administrationBolus permitted? Yes/no; individualized dose; explicit no-bolus choice if bleeding riskDo not apply systemic therapeutic-VTE nomograms to every extracorporeal circuit. [R5]
MaintenanceInfusion in IU/h or IU/kg/h and source (patient vs circuit)Adjust only against the designated validated local assay/target. [R1]
MonitoringaPTT or ACT or anti-Xa per procedure; time of first and repeat sample; platelet count and access bleedingaPTT targets depend on reagent and procedure; ACT and aPTT are not interchangeable. [R1] [R5]
Response / stopClotting, severe bleeding, HIT suspicion, unexplained thrombocytopenia or assay overshootSuspected HIT: stop all heparin and activate local HIT/non-heparin pathway.

5. Citrate: ACD-A vs regional citrate protocols

ACD-A in centrifugal apheresis

ACD-A is mixed with blood at the inlet of a centrifugal separator. Prescribe the approved whole-blood:ACD ratio and the separator's software-specific infusion mode. The patient receives some citrate with returned blood/plasma, particularly if plasma products are infused. Track systemic ionized calcium and symptoms. Do not assume ACD-A is acceptable to a downstream sorbent. [R1] [R5] [R9]

Citrate in membrane-based apheresis and adsorption circuits

Regional citrate anticoagulation may be considered only when the exact membrane-apheresis platform, anticoagulant formulation, calcium replacement and downstream cartridge are validated together. This is not the same as the ACD-A inlet regimen used in centrifugal apheresis. Prescribe the manufacturer-approved infusion sites, blood-flow linkage, sampling points and systemic ionized-calcium surveillance. No universal citrate rate, post-filter calcium target or calcium replacement rate can be inferred across TPE, DFPP, immunoadsorption or plasma adsorption.

If the complete circuit does not have an approved citrate protocol, do not extrapolate a regimen from a different extracorporeal technique. Escalate unexplained hypocalcemia, arrhythmia or acid–base disturbances and follow the local emergency protocol.

6. Therapeutic plasma exchange (TPE): centrifugal & membrane

Centrifugal TPE — practical protocol

  1. Select the separator's validated ACD-A regimen and document inlet whole-blood flow, citrate ratio, total planned plasma exchange and replacement fluid.
  2. Check baseline systemic iCa; evaluate added citrate load from donor plasma and impaired citrate metabolism. Plan calcium supplementation according to local protocol.
  3. Monitor citrate symptoms (perioral tingling, cramps, hypotension, dysrhythmia), iCa as indicated, vascular access and procedure-related coagulopathy.
  4. If systemic anticoagulants are already prescribed, review exposure and bleeding before and after TPE; TPE may remove anticoagulants and coagulation proteins. [R1] [R5]

Membrane TPE — practical protocol

  1. Check the membrane platform's permitted anticoagulants; select UFH or a specifically validated citrate configuration.
  2. Prescribe access, blood flow, filtration fraction, replacement fluid and anticoagulation delivery site; do not assume a centrifuge ACD ratio applies.
  3. Monitor filter TMP/pressure, visual clotting, platelets, fibrinogen, bleeding and the chosen anticoagulant-specific assays. [R1] [R5]
There is no single universally applicable TPE anticoagulation guideline; patient, replacement fluid and medications must be considered individually. [R5]

7. DFPP & cascade filtration

DFPP uses a plasma separator plus a secondary plasma fractionator. UFH is frequently used; RCA has been compared with systemic heparin in a retrospective kidney-transplant cohort. That study supports feasibility in its setting, not a universal RCA-DFPP dosing algorithm. [R4]

  1. Confirm anticoagulant compatibility with both filters, circuit tubing and the machine's fluid pathways.
  2. UFH pathway: prescribe local UFH bolus/infusion and validated aPTT/ACT target, with clinical bleeding checks.
  3. RCA pathway: use only an institution-validated DFPP citrate-calcium protocol; explicitly define sample location, target range and calcium replacement.
  4. Monitor pressures across primary and secondary filters, clotting, fibrinogen and coagulation tests. Replace depleted factors only as clinically indicated.
  5. Escalate if filter clotting, rising pressure, hypocalcemia or bleeding occurs. [R4] [R1]

8. Immunoadsorption & ABO-specific adsorption

Includes regenerative IA (e.g., immunoglobulin-binding columns), single-use IA and ABO-specific sorbents. Anticoagulation must cover the cell separator and the adsorption column; the plasma is generally returned rather than replaced. [R1] [R16]

  1. Identify whether separation is centrifugal (often ACD-A-based) or membrane-based; check column IFU for citrate/heparin compatibility and priming.
  2. Document exact column, regenerative/single-use setup, plasma flow, anticoagulant entry point and whether UFH supplementation is allowed.
  3. For ACD-A, prescribe separator-approved ratio and calcium monitoring; for UFH, prescribe local target and platelet/bleeding surveillance.
  4. Monitor access, clotting, pressure across adsorber and patient symptoms; review removal of therapeutic antibodies or drugs.
Evidence example, not a default dose: a Glycosorb ABO series used ACD-A together with sodium heparin at 1,000 IU/h. This single-center clinical protocol should not be generalized to all ABO or other IA devices. [R16]

9. Plasma adsorption, bilirubin adsorption & DPMAS

Includes plasma bilirubin adsorption, nonspecific resin adsorption, immune-complex/CRP adsorption, and DPMAS-like separated-plasma circuits. Anticoagulation applies to the entire blood–plasma–return circuit, not just the sorbent. For many proprietary columns, independent comparative anticoagulation studies are scarce; the manufacturer IFU and local validated procedure are primary. [R1] [R5]

  1. Map the circuit: plasma separator → adsorber(s) → return. Mark anticoagulant infusion and sampling sites.
  2. Check whether the specific sorbent permits heparin, citrate or both. Do not assume permission from a different model or brand.
  3. Document UFH monitoring or citrate-calcium protocol, planned processed plasma volume, filter/adsorber pressures and fluid balance.
  4. In liver dysfunction, be particularly cautious about citrate metabolism, hypocalcemia and acid–base abnormalities; monitor clinically and biochemically.
  5. Specify circuit-clotting and metabolic stop criteria.

10. Lipoprotein apheresis: LA-15, filtration, H.E.L.P. & direct adsorption

Kaneka LIPOSORBER LA-15 (US labeling): heparin-only. The manufacturer/FDA labeling explicitly states that citrate preparations (ACD) must never be used in the LA-15 system. Follow its approved heparin procedure, and monitor ACT/PTT as instructed. This warning does not establish identical restrictions for every other lipoprotein apheresis device. [R9] [R10]
TechniqueAnticoagulation prescription pointReference
Dextran-sulfate plasma adsorption (LA-15)Heparin-only per current US labeling; verify jurisdiction-specific IFU.[R9] [R10]
Other plasma LDL adsorption, MONET lipid filtrationUse exact device-approved anticoagulation; do not extrapolate LA-15 restrictions or ACD approval.[R19]
Direct whole-blood LDL adsorption (e.g., DALI)Different flow path; follow device-specific heparin/citrate instructions.[R19]
H.E.L.P. LDL precipitationHeparin is a required process reagent in the extracorporeal plasma precipitation stage; dedicated heparin handling and downstream removal are part of the validated system.[R11] [R20]

For all lipoprotein systems: monitor vascular access, bleeding, pressures, platelet count and potential removal of concomitant medications. [R19]

11. Whole-blood hemoadsorption: stand-alone and extracorporeal bypass

Stand-alone hemoperfusion

Do not assume that citrate is suitable for a stand-alone cartridge. CytoSorb specifically states heparin only for stand-alone therapy; citrate is contraindicated in that setup. Toraymyxin literature describes UFH or nafamostat in specific Japanese practice settings; neither is a generic regimen for Jafron, Biosky, Efferon, Seraph or other cartridges. [R7] [R17]

Hemoadsorption integrated with ECMO / cardiopulmonary bypass

Systemic anticoagulation is governed by the ECMO/CPB service's own protocol. A sorbent cannot be used as a reason to switch to regional citrate without a validated ECMO configuration. Record insertion location, flow and access pressures, and monitor bleeding/thrombosis. [R8]

Indication warning: the Surviving Sepsis Campaign 2026 suggests against blood purification (including hemoperfusion) as routine treatment for adult sepsis/septic shock and separately against polymyxin-B hemoperfusion. This does not automatically govern other indications. [R18]

12. Extracorporeal photopheresis (ECP)

ECP uses either an integrated photoactivation platform or a multistep leukapheresis/irradiation procedure. Anticoagulation options are system- and software-specific: ACD-A is common in multistep collection, while integrated THERAKOS systems have used heparin and (in current European materials) an ACD-A option. [R12] [R14] [R21]

ECP workflowPractical anticoagulation prescriptionSource
Multistep / off-line collectionUse collection separator's approved ACD-A ratio; monitor systemic iCa and symptoms; check the validated downstream product handling.[R12] [R21]
THERAKOS CELLEX with heparinFollow exact operator manual/software revision. One historical UK manual gives a heparinized saline setup and a default blood:anticoagulant ratio of 10:1; these are not universal ECP doses.[R13]
THERAKOS CELLEX with ACD-ACurrent European product information describes heparinized saline or ACD-A choice; verify region, software and consumables.[R14] [R15]
A retrospective real-world device-data review of 23,334 ECP sessions described both heparin and ACD-A use; observational device performance data are not evidence that the two can be substituted without validation. [R15]

13. Safety monitoring, troubleshooting & stop rules

ProblemImmediate assessmentProtocol response
Visible clot / rising TMP / adsorber pressureCheck access, kinks, flow, filter/column pressure, anticoagulant delivery and sampling validity.Pause or stop as indicated; never automatically increase anticoagulation without assessing bleeding and device limits.
Bleeding / marked aPTT/ACT prolongationCheck UFH delivery, platelet count, coagulation, fibrinogen (TPE/DFPP), other anticoagulants.Activate local bleeding/UFH reversal protocol; consider stopping therapy and clinical escalation.
Suspected HITNew thrombocytopenia/thrombosis with heparin exposure.Stop all heparin; use specialist HIT pathway and validated alternative anticoagulation.
Citrate symptoms / low systemic iCaAssess paresthesia, cramps, hypotension, ECG, ionized calcium and citrate input.Follow calcium correction and citrate adjustment/stop algorithm; urgent escalation for severe symptoms. [R1]
Suspected citrate accumulationRising total/iCa ratio (often ≥2.5), increasing calcium needs, acid–base derangement, poor perfusion.Escalate urgently; reduce/stop citrate as directed by validated protocol and correct abnormalities.
Hypersensitivity / bradykinin reactionHypotension, rash, respiratory symptoms, ACE inhibitor exposure with relevant sorbents.Stop and manage as emergency; review specific column contraindications. [R10]

14. Calculation aid: documentation, not automatic dosing

These calculations are arithmetic checks only. They do not determine whether a regimen is appropriate or compatible with the device.

21.60 LCalculated blood processed

—Calculated total infusion (excludes bolus)

—Total/iCa ratio (same units required)

Never use a ratio alone to diagnose citrate accumulation; integrate systemic iCa, calcium needs, acid–base status and patient context. These calculators do not prescribe heparin bolus, citrate rate, calcium infusion or a treatment duration.

15. Prescription Worksheet — complete & print

Complete locally for the actual patient and device. No values are preauthorized. The page does not transmit or store patient information.

Mandatory validation: the completed worksheet is a draft until the responsible clinical team verifies all numeric orders, monitoring, and compatibility with the current local IFU and institutional anticoagulation protocol.

16. References & evidence provenance

Each section links to the supporting reference numbers below. Sources range from clinical guidelines and trials to manufacturer-specific IFUs; a manufacturer's device instruction has precedence for compatibility and permissible anticoagulants. Evidence from one device is not automatically generalizable to another.

  1. R1. Lee G, Arepally GM. Anticoagulation techniques in apheresis: from heparin to citrate and beyond. J Clin Apher. 2012;27:117–125. Read source ↗
  2. R4. Teh SP et al. Regional citrate vs systemic heparin for DFPP in kidney transplant recipients. J Clin Apher. 2023;38:16–23. Read source ↗
  3. R5. The management of anticoagulation in patients undergoing therapeutic plasma exchange: a concise review. J Clin Apher. 2018. Read source ↗
  4. R7. CytoSorbents. Anticoagulation FAQ; distinguish stand-alone hemoperfusion from combined-circuit therapy. Read source ↗
  5. R8. CytoSorb. Technical Data / Application Guide (manufacturer; 2024 revision). Read source ↗
  6. R9. Kaneka. LIPOSORBER LA-15 US safety information / manufacturer instructions: ACD not permitted in this system. Read source ↗
  7. R10. US FDA. LIPOSORBER LA-15 System Labeling (manufacturer-specific heparin-only anticoagulation). Read source ↗
  8. R11. B. Braun. H.E.L.P. Plasmat Futura brief instructions for use; heparin is a process reagent and anticoagulation requires system-specific procedures. Read source ↗
  9. R12. European Dermatology Forum. Updated ECP guidelines, 2020, Part 1. Read source ↗
  10. R13. THERAKOS CELLEX Operator Manual (UK edition; device-specific heparin setup and A/C ratio). Read source ↗
  11. R14. THERAKOS CELLEX current European product information: device-specific option of ACD-A or heparinized saline. Read source ↗
  12. R15. Real-world evidence of heparin and citrate use in ECP: device data from 23,334 sessions. J Clin Apher. 2024. Read source ↗
  13. R16. Glycosorb ABO immunoadsorption clinical series: combined ACD-A and heparin example, not a general prescription. Read source ↗
  14. R17. Toraymyxin direct hemoperfusion review, including anticoagulation regimens and Japan-specific nafamostat use. Read source ↗
  15. R18. Prescott H et al. Surviving Sepsis Campaign: International Guidelines for Sepsis and Septic Shock 2026. Read source ↗
  16. R19. Feingold KR. Lipoprotein Apheresis. Endotext. Updated February 2026. Read source ↗
  17. R20. B. Braun. H.E.L.P. apheresis process and Plasmat Futura platform. Read source ↗
  18. R21. Extracorporeal photopheresis in graft-versus-host disease: inline and offline anticoagulation comparison. 2020. Read source ↗
Evidence hierarchy: current IFU and regulatory labeling → institution-approved protocol → technique-specific guidelines and randomized evidence → cohort/observational evidence → manufacturer technical guides. The 2026 KDIGO update is not yet a finalized guideline; the 2012 KDIGO guideline remains the published reference at this date.