1. Scope, terminology & safety
Anticoagulation is a circuit decision, not simply a disease decision. Determine whether blood is centrifuged, filtered through a membrane, passed through a plasma sorbent, or perfused through a whole-blood cartridge. Heparin and citrate have distinct systemic and regional effects. A patient already receiving systemic anticoagulation may need a different extracorporeal strategy. [R1] [R5]
2. Technique-by-technique anticoagulation matrix
“Possible” means reported or used in validated configurations; it does not imply approval for every machine, filter, country or indication.
| Technique / circuit | Heparin (UFH) | Citrate | Protocol decision & evidence |
|---|---|---|---|
| Centrifugal TPE | Alternative or adjunct in selected patients | Common default: ACD-A | Use separator-approved blood:ACD ratio; consider plasma replacement citrate load and calcium. [R1] [R5] |
| Membrane TPE | Common option | Possible only with validated membrane/RCA setup | Check filter, replacement fluid, bleeding risk, citrate-calcium balance. [R1] [R5] |
| DFPP / cascade filtration | Common comparator | RCA studied; not universal | Two filters, pressure/clotting and fibrinogen loss require separate checks. [R4] |
| Immunoadsorption (IA), incl. ABO | Alone or in approved combination | Often ACD-A at centrifugal separation | Separator and adsorber must both permit chosen anticoagulant; a published ABO series used combined therapy. [R1] [R16] |
| Plasma adsorption: bilirubin, liver support, DPMAS | Commonly described | Only if complete circuit validated | Plasma separator + adsorber + return line; no cross-brand universal dose. [R1] [R5] |
| LDL / Lp(a) plasma adsorption | Depends on device; required in LA-15 US system | Prohibited in LA-15 US system; other systems differ | Check exact lipoprotein device; do not generalize. [R9] [R10] [R19] |
| MONET / plasma lipid filtration | Device-dependent | Device-dependent | Plasma filter circuit; no generic dose or target extrapolated from DFPP. [R19] |
| H.E.L.P. LDL precipitation | Process-specific heparin and system-specific anticoagulation | Do not substitute without validated IFU | Heparin is part of the precipitation chemistry; not merely an optional anticoagulant. [R11] [R20] |
| Whole-blood hemoadsorption — stand-alone | Common; device-specific | Not automatically permitted | CytoSorb stand-alone: manufacturer says heparin only. Other cartridges have different IFUs. [R7] [R17] |
| Hemoadsorption with ECMO / CPB | Often governed by primary ECMO/CPB anticoagulation protocol | Not a drop-in replacement for systemic ECMO anticoagulation | Dedicated ECMO/CPB team sets systemic targets; sorbent IFU must allow placement. [R7] [R8] |
| ECP (photopheresis) | Some integrated systems | ACD-A in many off-line circuits; option in current CELLEX EU materials | Model/software-specific. Do not transplant historical CELLEX heparin doses to other ECP systems. [R12] [R13] [R14] [R15] |
3. Before starting: prescription & risk screen
- Confirm clinical indication and whether the proposed extracorporeal treatment itself is justified; anticoagulation does not establish treatment efficacy. For adult sepsis/septic shock, the 2026 Surviving Sepsis Campaign suggests against routine blood purification and polymyxin-B hemoperfusion. [R18]
- Record patient weight, active bleeding, surgery/procedures, history of HIT, prior citrate intolerance, shock/hypoperfusion, severe hepatic dysfunction, systemic anticoagulants and antiplatelets.
- Record baseline CBC/platelets, PT/INR, aPTT (and anti-Xa if indicated), fibrinogen for plasma-removing therapies, systemic ionized calcium, magnesium, sodium, bicarbonate/ABG and lactate if RCA considered. [R1] [R5]
- Document exact device model, software, circuit and disposable IFU revision, access, priming, blood/plasma flows, fluid composition, replacement and circuit volume.
- Choose anticoagulation: ACD-A, UFH, RCA, anticoagulant-free if appropriate and validated, or specialist non-heparin strategy for HIT. Document reason and monitoring plan. [R1]
- Assign clinician to approve start, monitor abnormalities, adjust infusion, and stop therapy if safety thresholds are breached.
4. UFH: prescription and monitoring protocol
| Order element | What to prescribe | Safety / evidence |
|---|---|---|
| Preparation | UFH concentration (IU/mL), line and priming instructions, separate systemic infusion vs circuit injection | Never infer priming concentration from a different cartridge. [R1] [R13] |
| Initial administration | Bolus permitted? Yes/no; individualized dose; explicit no-bolus choice if bleeding risk | Do not apply systemic therapeutic-VTE nomograms to every extracorporeal circuit. [R5] |
| Maintenance | Infusion in IU/h or IU/kg/h and source (patient vs circuit) | Adjust only against the designated validated local assay/target. [R1] |
| Monitoring | aPTT or ACT or anti-Xa per procedure; time of first and repeat sample; platelet count and access bleeding | aPTT targets depend on reagent and procedure; ACT and aPTT are not interchangeable. [R1] [R5] |
| Response / stop | Clotting, severe bleeding, HIT suspicion, unexplained thrombocytopenia or assay overshoot | Suspected HIT: stop all heparin and activate local HIT/non-heparin pathway. |
5. Citrate: ACD-A vs regional citrate protocols
ACD-A in centrifugal apheresis
ACD-A is mixed with blood at the inlet of a centrifugal separator. Prescribe the approved whole-blood:ACD ratio and the separator's software-specific infusion mode. The patient receives some citrate with returned blood/plasma, particularly if plasma products are infused. Track systemic ionized calcium and symptoms. Do not assume ACD-A is acceptable to a downstream sorbent. [R1] [R5] [R9]
Citrate in membrane-based apheresis and adsorption circuits
Regional citrate anticoagulation may be considered only when the exact membrane-apheresis platform, anticoagulant formulation, calcium replacement and downstream cartridge are validated together. This is not the same as the ACD-A inlet regimen used in centrifugal apheresis. Prescribe the manufacturer-approved infusion sites, blood-flow linkage, sampling points and systemic ionized-calcium surveillance. No universal citrate rate, post-filter calcium target or calcium replacement rate can be inferred across TPE, DFPP, immunoadsorption or plasma adsorption.
6. Therapeutic plasma exchange (TPE): centrifugal & membrane
Centrifugal TPE — practical protocol
- Select the separator's validated ACD-A regimen and document inlet whole-blood flow, citrate ratio, total planned plasma exchange and replacement fluid.
- Check baseline systemic iCa; evaluate added citrate load from donor plasma and impaired citrate metabolism. Plan calcium supplementation according to local protocol.
- Monitor citrate symptoms (perioral tingling, cramps, hypotension, dysrhythmia), iCa as indicated, vascular access and procedure-related coagulopathy.
- If systemic anticoagulants are already prescribed, review exposure and bleeding before and after TPE; TPE may remove anticoagulants and coagulation proteins. [R1] [R5]
Membrane TPE — practical protocol
- Check the membrane platform's permitted anticoagulants; select UFH or a specifically validated citrate configuration.
- Prescribe access, blood flow, filtration fraction, replacement fluid and anticoagulation delivery site; do not assume a centrifuge ACD ratio applies.
- Monitor filter TMP/pressure, visual clotting, platelets, fibrinogen, bleeding and the chosen anticoagulant-specific assays. [R1] [R5]
7. DFPP & cascade filtration
DFPP uses a plasma separator plus a secondary plasma fractionator. UFH is frequently used; RCA has been compared with systemic heparin in a retrospective kidney-transplant cohort. That study supports feasibility in its setting, not a universal RCA-DFPP dosing algorithm. [R4]
- Confirm anticoagulant compatibility with both filters, circuit tubing and the machine's fluid pathways.
- UFH pathway: prescribe local UFH bolus/infusion and validated aPTT/ACT target, with clinical bleeding checks.
- RCA pathway: use only an institution-validated DFPP citrate-calcium protocol; explicitly define sample location, target range and calcium replacement.
- Monitor pressures across primary and secondary filters, clotting, fibrinogen and coagulation tests. Replace depleted factors only as clinically indicated.
- Escalate if filter clotting, rising pressure, hypocalcemia or bleeding occurs. [R4] [R1]
8. Immunoadsorption & ABO-specific adsorption
Includes regenerative IA (e.g., immunoglobulin-binding columns), single-use IA and ABO-specific sorbents. Anticoagulation must cover the cell separator and the adsorption column; the plasma is generally returned rather than replaced. [R1] [R16]
- Identify whether separation is centrifugal (often ACD-A-based) or membrane-based; check column IFU for citrate/heparin compatibility and priming.
- Document exact column, regenerative/single-use setup, plasma flow, anticoagulant entry point and whether UFH supplementation is allowed.
- For ACD-A, prescribe separator-approved ratio and calcium monitoring; for UFH, prescribe local target and platelet/bleeding surveillance.
- Monitor access, clotting, pressure across adsorber and patient symptoms; review removal of therapeutic antibodies or drugs.
9. Plasma adsorption, bilirubin adsorption & DPMAS
Includes plasma bilirubin adsorption, nonspecific resin adsorption, immune-complex/CRP adsorption, and DPMAS-like separated-plasma circuits. Anticoagulation applies to the entire blood–plasma–return circuit, not just the sorbent. For many proprietary columns, independent comparative anticoagulation studies are scarce; the manufacturer IFU and local validated procedure are primary. [R1] [R5]
- Map the circuit: plasma separator → adsorber(s) → return. Mark anticoagulant infusion and sampling sites.
- Check whether the specific sorbent permits heparin, citrate or both. Do not assume permission from a different model or brand.
- Document UFH monitoring or citrate-calcium protocol, planned processed plasma volume, filter/adsorber pressures and fluid balance.
- In liver dysfunction, be particularly cautious about citrate metabolism, hypocalcemia and acid–base abnormalities; monitor clinically and biochemically.
- Specify circuit-clotting and metabolic stop criteria.
10. Lipoprotein apheresis: LA-15, filtration, H.E.L.P. & direct adsorption
| Technique | Anticoagulation prescription point | Reference |
|---|---|---|
| Dextran-sulfate plasma adsorption (LA-15) | Heparin-only per current US labeling; verify jurisdiction-specific IFU. | [R9] [R10] |
| Other plasma LDL adsorption, MONET lipid filtration | Use exact device-approved anticoagulation; do not extrapolate LA-15 restrictions or ACD approval. | [R19] |
| Direct whole-blood LDL adsorption (e.g., DALI) | Different flow path; follow device-specific heparin/citrate instructions. | [R19] |
| H.E.L.P. LDL precipitation | Heparin is a required process reagent in the extracorporeal plasma precipitation stage; dedicated heparin handling and downstream removal are part of the validated system. | [R11] [R20] |
For all lipoprotein systems: monitor vascular access, bleeding, pressures, platelet count and potential removal of concomitant medications. [R19]
11. Whole-blood hemoadsorption: stand-alone and extracorporeal bypass
Stand-alone hemoperfusion
Do not assume that citrate is suitable for a stand-alone cartridge. CytoSorb specifically states heparin only for stand-alone therapy; citrate is contraindicated in that setup. Toraymyxin literature describes UFH or nafamostat in specific Japanese practice settings; neither is a generic regimen for Jafron, Biosky, Efferon, Seraph or other cartridges. [R7] [R17]
Hemoadsorption integrated with ECMO / cardiopulmonary bypass
Systemic anticoagulation is governed by the ECMO/CPB service's own protocol. A sorbent cannot be used as a reason to switch to regional citrate without a validated ECMO configuration. Record insertion location, flow and access pressures, and monitor bleeding/thrombosis. [R8]
12. Extracorporeal photopheresis (ECP)
ECP uses either an integrated photoactivation platform or a multistep leukapheresis/irradiation procedure. Anticoagulation options are system- and software-specific: ACD-A is common in multistep collection, while integrated THERAKOS systems have used heparin and (in current European materials) an ACD-A option. [R12] [R14] [R21]
| ECP workflow | Practical anticoagulation prescription | Source |
|---|---|---|
| Multistep / off-line collection | Use collection separator's approved ACD-A ratio; monitor systemic iCa and symptoms; check the validated downstream product handling. | [R12] [R21] |
| THERAKOS CELLEX with heparin | Follow exact operator manual/software revision. One historical UK manual gives a heparinized saline setup and a default blood:anticoagulant ratio of 10:1; these are not universal ECP doses. | [R13] |
| THERAKOS CELLEX with ACD-A | Current European product information describes heparinized saline or ACD-A choice; verify region, software and consumables. | [R14] [R15] |
13. Safety monitoring, troubleshooting & stop rules
| Problem | Immediate assessment | Protocol response |
|---|---|---|
| Visible clot / rising TMP / adsorber pressure | Check access, kinks, flow, filter/column pressure, anticoagulant delivery and sampling validity. | Pause or stop as indicated; never automatically increase anticoagulation without assessing bleeding and device limits. |
| Bleeding / marked aPTT/ACT prolongation | Check UFH delivery, platelet count, coagulation, fibrinogen (TPE/DFPP), other anticoagulants. | Activate local bleeding/UFH reversal protocol; consider stopping therapy and clinical escalation. |
| Suspected HIT | New thrombocytopenia/thrombosis with heparin exposure. | Stop all heparin; use specialist HIT pathway and validated alternative anticoagulation. |
| Citrate symptoms / low systemic iCa | Assess paresthesia, cramps, hypotension, ECG, ionized calcium and citrate input. | Follow calcium correction and citrate adjustment/stop algorithm; urgent escalation for severe symptoms. [R1] |
| Suspected citrate accumulation | Rising total/iCa ratio (often ≥2.5), increasing calcium needs, acid–base derangement, poor perfusion. | Escalate urgently; reduce/stop citrate as directed by validated protocol and correct abnormalities. |
| Hypersensitivity / bradykinin reaction | Hypotension, rash, respiratory symptoms, ACE inhibitor exposure with relevant sorbents. | Stop and manage as emergency; review specific column contraindications. [R10] |
14. Calculation aid: documentation, not automatic dosing
These calculations are arithmetic checks only. They do not determine whether a regimen is appropriate or compatible with the device.
21.60 LCalculated blood processed
—Calculated total infusion (excludes bolus)
—Total/iCa ratio (same units required)
Never use a ratio alone to diagnose citrate accumulation; integrate systemic iCa, calcium needs, acid–base status and patient context. These calculators do not prescribe heparin bolus, citrate rate, calcium infusion or a treatment duration.
15. Prescription Worksheet — complete & print
Complete locally for the actual patient and device. No values are preauthorized. The page does not transmit or store patient information.
16. References & evidence provenance
Each section links to the supporting reference numbers below. Sources range from clinical guidelines and trials to manufacturer-specific IFUs; a manufacturer's device instruction has precedence for compatibility and permissible anticoagulants. Evidence from one device is not automatically generalizable to another.
- R1. Lee G, Arepally GM. Anticoagulation techniques in apheresis: from heparin to citrate and beyond. J Clin Apher. 2012;27:117–125. Read source ↗
- R4. Teh SP et al. Regional citrate vs systemic heparin for DFPP in kidney transplant recipients. J Clin Apher. 2023;38:16–23. Read source ↗
- R5. The management of anticoagulation in patients undergoing therapeutic plasma exchange: a concise review. J Clin Apher. 2018. Read source ↗
- R7. CytoSorbents. Anticoagulation FAQ; distinguish stand-alone hemoperfusion from combined-circuit therapy. Read source ↗
- R8. CytoSorb. Technical Data / Application Guide (manufacturer; 2024 revision). Read source ↗
- R9. Kaneka. LIPOSORBER LA-15 US safety information / manufacturer instructions: ACD not permitted in this system. Read source ↗
- R10. US FDA. LIPOSORBER LA-15 System Labeling (manufacturer-specific heparin-only anticoagulation). Read source ↗
- R11. B. Braun. H.E.L.P. Plasmat Futura brief instructions for use; heparin is a process reagent and anticoagulation requires system-specific procedures. Read source ↗
- R12. European Dermatology Forum. Updated ECP guidelines, 2020, Part 1. Read source ↗
- R13. THERAKOS CELLEX Operator Manual (UK edition; device-specific heparin setup and A/C ratio). Read source ↗
- R14. THERAKOS CELLEX current European product information: device-specific option of ACD-A or heparinized saline. Read source ↗
- R15. Real-world evidence of heparin and citrate use in ECP: device data from 23,334 sessions. J Clin Apher. 2024. Read source ↗
- R16. Glycosorb ABO immunoadsorption clinical series: combined ACD-A and heparin example, not a general prescription. Read source ↗
- R17. Toraymyxin direct hemoperfusion review, including anticoagulation regimens and Japan-specific nafamostat use. Read source ↗
- R18. Prescott H et al. Surviving Sepsis Campaign: International Guidelines for Sepsis and Septic Shock 2026. Read source ↗
- R19. Feingold KR. Lipoprotein Apheresis. Endotext. Updated February 2026. Read source ↗
- R20. B. Braun. H.E.L.P. apheresis process and Plasmat Futura platform. Read source ↗
- R21. Extracorporeal photopheresis in graft-versus-host disease: inline and offline anticoagulation comparison. 2020. Read source ↗